Analog input is not available in Indigo510
transmitters or transmitters that are powered with Power over Ethernet (PoE).
Only 1 measurement probe can be
connected to the transmitter when you power the analog input device through Indigo520 probe 2
connection terminal. The measurement probe has to be connected to probe 1 connection
terminal.
The Indigo520 transmitter has 1 analog
current (mA) input.
Before you start, make sure that the probes
you want to use are connected to the transmitter.
Figure 1. Analog input configuration menu in web interface
Connect to the web interface.
Log in to the web interface as an administrator user.
Select Inputs and outputs > Analog input.
In Analog input
off/on, set analog input on.
To power your analog input device through Indigo520, in Analog input
powering, set analog input powering on.