Analog output terminals and lead-through - INDIGO500SERIES - INDIGO510 - INDIGO520

Indigo510, Indigo520 User Guide

Document code
M212287EN
Revision
P
Language
English
Product
INDIGO500SERIES
INDIGO510
INDIGO520
Document type
User guide

Before connecting wires or cables, make sure that the transmitter is powered off.

Analog outputs are not available in transmitters that are powered with Power over Ethernet (PoE).

Use the touchscreen or web interface to change the output mode (for example, 0 … 5 V or 4 … 20 mA) and scaling of the analog outputs.

For the M20×1.5 cable gland ordered together with the transmitter from Vaisala, the cable diameter is 5.0 … 8.0 mm (0.20 … 0.31 in). Tightening torque for the cable gland is 8 Nm.

For the M20×1.5 cable gland with split bushing, the cable diameter is 7 mm (0.28 in).

Table 1. Analog output terminals
Terminal Function Notes
CH1 + Analog output channel 1 + Max. wire size: 2.5 mm2 (14 AWG)
CH1 - Analog output channel 1 -
CH2 + Analog output channel 2 +
CH2 - Analog output channel 2 -
CH3 + Analog output channel 3 +
CH3 - Analog output channel 3 -
CH4 + Analog output channel 4 +