Indigo510 specifications - INDIGO500SERIES - INDIGO510 - INDIGO520

Indigo510, Indigo520 User Guide

Document code
M212287EN
Revision
P
Language
English
Product
INDIGO500SERIES
INDIGO510
INDIGO520
Document type
User guide
Table 1. Indigo510 transmitter options
Property Description/Value
Display
  • Capacitive touchscreen display
  • No display (indicator LED) 1
Powering Protective extra-low voltage (11 … 35 V DC, 24 V AC ± 15% 50/60 Hz)
Table 2. Indigo510 operating environment
Property Description/Value
For use in wet locations Yes
Operating humidity

0 … 100 %RH

Maximum operating altitude

4000 m (13 123 ft)

IP rating

IP66: Dust-tight. Protected from powerful water jets from any direction. 2

Operating temperature
With display −20 … +60 °C (−4 … +140 °F)
Without display −40 … +60 °C (−40 … +140 °F)
Storage temperature
With display

−30 … +60 °C (−22 … +158 °F)

Without display

−40 … +60 °C (−40 … +140 °F)

Table 3. Indigo510 inputs and outputs
Property Description/Value
Operating power
Protective extra-low voltage (PELV)

11 … 35 V DC, 24 V AC ±15 % 50/60 Hz, max. current 2 A (power supply is galvanically isolated)

Fuse size for power supply: 3 A

Isolation voltage: 500 V AC, 1000 V DC

PELV power cable temp. rating ≥ +80 °C (+176 °F)
Typical current consumption at +20 °C (+68 °F) (Uin 24 V DC) 3
Base consumption (no display, analog outputs, or communication) 50 mA
With display + 60 mA
With voltage analog output < 2 mA per channel
With current analog output + 21 mA per channel
With Ethernet cable connected + 15 mA
Analog outputs
Number of analog outputs 2
Isolation Isolated from power supply
Selectable voltage output types 0 … 1 V, 0 … 5 V, 0 … 10 V, scalable
Selectable current output types 4 … 20 mA, 0 … 20 mA, scalable
Max. wire size 2.5 mm2 (14 AWG)
Accuracy of analog outputs at +20 °C (+68 °F) ±0.05 % full scale
Temperature dependence ±0.005 % / °C full scale
External loads:
Current outputs RL < 500 Ω
0 … 1 V output RL > 2 kΩ
0 … 5 V and 0 … 10 V outputs RL > 10 kΩ
Ethernet interface
Supported standards 10BASE-T, 100BASE-TX
Connector 8P8C (RJ45)
Supported protocols Modbus TCP/IP (port 502), HTTPS (port 8443)
Table 4. Indigo510 compliance
Property Description/Value

EU directives and regulations

EMC Directive (2014/30/EU)

RoHS Directive (2011/65/EU) amended by 2015/863

Electromagnetic compatibility (EMC)

IEC/EN 61326‑1, industrial environment

CISPR 32 / EN 55032, Class B

Electrical safety

IEC/EN 61010‑1

Type approvals

DNV GL certificate no. TAA000032M

Compliance marks

CE, China RoHS, FCC, RCM, UKCA

Listing marks

UL Listed (USA and Canada)

FCC compliance

FCC Part 15, Class B

Table 5. Indigo500 Series mechanical specifications
Property Description/Value

UL 50E (NEMA) rating

NEMA 4: Hose down and splashing water, NEMA 250-2014 Type 4

Housing classification IK08, DIN EN ISO 11997-1: Cycle B (VDA 621-415)
Housing material AlSi10Mg (DIN 1725)
Display window material Strengthened glass (IK08)
Weight 1.5 kg (3.3 lb)
Dimensions (H × W × D) 142 × 182 × 67 mm (5.63 × 7.17 × 2.64 in)
Cable diameters for cable glands
M20×1.5 glands 5.0 … 9.0 mm (0.20 … 0.31 in)
M20×1.5 glands with split bushing 7 mm (0.28 in)
M16×1.5 glands 2.0 … 6.0 mm (0.08 … 0.24 in)
Table 6. Indigo500 Series user interfaces
Property Description/Value
User interfaces Web interface for remote use, optional touchscreen display
Supported languages English, Chinese, French, German, Japanese, Spanish
Optional display 5" capacitive touchscreen
Integrated data logging capabilities Non-volatile memory, at least 10 years' storage with 24 h interval logging
Table 7. Indigo-compatible probes
Measurement type Probe models
Humidity and temperature HMP1, HMP3, HMP4, HMP5, HMP7, HMP8, HMP9
Temperature TMP1
Dew point DMP5, DMP6, DMP7, DMP8
Carbon dioxide GMP251, GMP252
Vaporized hydrogen peroxide HPP271, HPP272
Moisture in oil MMP8
Table 8. Other Indigo510 compatible devices
Device or series Models
MHT410 Moisture, Hydrogen and Temperature Transmitter MHT410
Indigo80 Handheld Indicator Indigo80
1 Recommended when the transmitter is exposed to direct UV light, and for outdoor installations and high-humidity environments.
2 Evaluated by Eurofins, not by UL.
3 For the current consumption of the connected measurement device, see the device's documentation, available at docs.vaisala.com.