Property | Description/Value |
---|---|
Display |
|
Powering | Protective extra-low voltage (11 … 35 V DC, 24 V AC ± 15% 50/60 Hz) |
Property | Description/Value |
---|---|
For use in wet locations | Yes |
Operating humidity |
0 … 100 %RH |
Maximum operating altitude |
4000 m (13 123 ft) |
IP rating |
IP66: Dust-tight. Protected from powerful water jets from any direction. 2 |
Operating temperature | |
With display | −20 … +60 °C (−4 … +140 °F) |
Without display | −40 … +60 °C (−40 … +140 °F) |
Storage temperature | |
With display |
−30 … +60 °C (−22 … +158 °F) |
Without display |
−40 … +60 °C (−40 … +140 °F) |
Property | Description/Value |
---|---|
Operating power | |
Protective extra-low voltage (PELV) |
11 … 35 V DC, 24 V AC ±15 % 50/60 Hz, max. current 2 A (power supply is galvanically isolated) Fuse size for power supply: 3 A Isolation voltage: 500 V AC, 1000 V DC |
PELV power cable temp. rating | ≥ +80 °C (+176 °F) |
Typical current consumption at +20 °C (+68 °F) (Uin 24 V DC) 3 | |
Base consumption (no display, analog outputs, or communication) | 50 mA |
With display | + 60 mA |
With voltage analog output | < 2 mA per channel |
With current analog output | + 21 mA per channel |
With Ethernet cable connected | + 15 mA |
Analog outputs | |
Number of analog outputs | 2 |
Isolation | Isolated from power supply |
Selectable voltage output types | 0 … 1 V, 0 … 5 V, 0 … 10 V, scalable |
Selectable current output types | 4 … 20 mA, 0 … 20 mA, scalable |
Max. wire size | 2.5 mm2 (14 AWG) |
Accuracy of analog outputs at +20 °C (+68 °F) | ±0.05 % full scale |
Temperature dependence | ±0.005 % / °C full scale |
External loads: | |
Current outputs | RL < 500 Ω |
0 … 1 V output | RL > 2 kΩ |
0 … 5 V and 0 … 10 V outputs | RL > 10 kΩ |
Ethernet interface | |
Supported standards | 10BASE-T, 100BASE-TX |
Connector | 8P8C (RJ45) |
Supported protocols | Modbus TCP/IP (port 502), HTTPS (port 8443) |
Property | Description/Value |
---|---|
EU directives and regulations |
EMC Directive (2014/30/EU) RoHS Directive (2011/65/EU) amended by 2015/863 |
Electromagnetic compatibility (EMC) |
IEC/EN 61326‑1, industrial environment CISPR 32 / EN 55032, Class B |
Electrical safety |
IEC/EN 61010‑1 |
Type approvals |
DNV GL certificate no. TAA000032M |
Compliance marks |
CE, China RoHS, FCC, RCM, UKCA |
Listing marks |
UL Listed (USA and Canada) |
FCC compliance |
FCC Part 15, Class B |
Property | Description/Value |
---|---|
UL 50E (NEMA) rating |
NEMA 4: Hose down and splashing water, NEMA 250-2014 Type 4 |
Housing classification | IK08, DIN EN ISO 11997-1: Cycle B (VDA 621-415) |
Housing material | AlSi10Mg (DIN 1725) |
Display window material | Strengthened glass (IK08) |
Weight | 1.5 kg (3.3 lb) |
Dimensions (H × W × D) | 142 × 182 × 67 mm (5.63 × 7.17 × 2.64 in) |
Cable diameters for cable glands | |
M20×1.5 glands | 5.0 … 9.0 mm (0.20 … 0.31 in) |
M20×1.5 glands with split bushing | 7 mm (0.28 in) |
M16×1.5 glands | 2.0 … 6.0 mm (0.08 … 0.24 in) |
Property | Description/Value |
---|---|
User interfaces | Web interface for remote use, optional touchscreen display |
Supported languages | English, Chinese, French, German, Japanese, Spanish |
Optional display | 5" capacitive touchscreen |
Integrated data logging capabilities | Non-volatile memory, at least 10 years' storage with 24 h interval logging |
Measurement type | Probe models |
---|---|
Humidity and temperature | HMP1, HMP3, HMP4, HMP5, HMP7, HMP8, HMP9 |
Temperature | TMP1 |
Dew point | DMP5, DMP6, DMP7, DMP8 |
Carbon dioxide | GMP251, GMP252 |
Vaporized hydrogen peroxide | HPP271, HPP272 |
Moisture in oil | MMP8 |
Device or series | Models |
---|---|
MHT410 Moisture, Hydrogen and Temperature Transmitter | MHT410 |
Indigo80 Handheld Indicator | Indigo80 |