Property | Description/Value |
---|---|
Display |
|
Powering |
|
Property | Description/Value |
---|---|
Barometric pressure (optional module) | |
Pressure range | 500 … 1100 hPa |
Class A: | |
Linearity | ±0.05 hPa |
Hysteresis | ±0.03 hPa |
Repeatability | ±0.03 hPa |
Calibration uncertainty | ±0.07 hPa |
Accuracy at +20 °C / +68 °F | ±0.10 hPa |
Temperature dependence | ±0.1 hPa |
Total accuracy (-40 … +60 °C / -40 … +140 °F) | ±0.15 hPa |
Long-term stability/year | ±0.1 hPa |
Response time (100 % response): | |
One sensor | 2 s |
Pressure units | hPa, mbar, kPa, Pa, inHg, mmH20, mmHg, torr, psia |
Property | Description/Value |
---|---|
For use in wet locations | Yes |
Operating humidity |
0 … 100 %RH |
Maximum operating altitude, AC (mains) power |
3000 m (9843 ft) |
Maximum operating altitude, protective extra-low voltage (PELV) and Power over Ethernet (PoE) | 4000 m (13 123 ft) |
IP rating |
IP66: Dust-tight. Protected from powerful water jets from any direction. 2 |
Operating temperature | |
With display | −20 … +55 °C (−4 … +131 °F) |
Without display | −40 … +60 °C (−40 … +140 °F) |
Without display with barometer module | −40 … +55 °C (−40 … +131 °F) |
Storage temperature | |
With display |
−30 … +60 °C (−22 … +158 °F) |
Without display |
−40 … +60 °C (−40 … +140 °F) |
Property | Description/Value |
---|---|
Operating power 3 | |
Protective extra-low voltage (PELV) version | 15 … 35 V DC, 24 V AC ±20 % 50/60 Hz, max. current 2 A (power supply is
galvanically isolated) Fuse size for power supply: 3 A Isolation voltage: 500 V AC, 1000 V DC |
PELV power cable temp. rating | ≥ +80 °C (+176 °F) |
AC (mains) power version |
100 … 240 V AC 50/60 Hz, max. current 1 A (power supply is galvanically isolated) Fuse size for power supply: 10 A Isolation voltage: 1500 V AC |
Power over Ethernet version 4 |
Power over Ethernet (PoE) IEEE 802.3at Type 2 Class 4 Max. current 600 mA, max. power consumption 25.5 W Isolation voltage: 500 V AC, 1000 V DC |
Typical current consumption at +20 °C (+68 °F) (Uin 24 V DC) 5 | |
Base consumption (no display, analog outputs, or communication) | 50 mA |
With display | + 60 mA |
With voltage analog output | < 2 mA per channel |
With current analog output | + 21 mA per channel |
With relays | + 9 mA per relay |
With Ethernet cable connected | + 15 mA |
With barometer module | + 5 mA |
Analog input | |
Available ranges | 4 ... 20 mA |
Resolution | 6 μA |
Display resolution | 0.01 mA |
Accuracy | ±0.05 mA |
Input impedances | 200 Ω |
Isolation | Isolated from power supply |
Overload protection | 40 mA max. (reverse polarity protected) |
Analog outputs | |
Number of analog outputs | 4 |
Isolation | Isolated from power supply |
Selectable voltage output types | 0 … 1 V, 0 … 5 V, 0 … 10 V, scalable |
Selectable current output types | 4 … 20 mA, 0 … 20 mA, scalable |
Max. wire size | 2.5 mm2 (14 AWG) |
Accuracy of analog outputs at +20 °C (+68 °F) | ±0.05 % full scale |
Temperature dependence | ±0.005 % / °C full scale |
External loads: | |
Current outputs | RL < 500 Ω |
0 … 1 V output | RL > 2 kΩ |
0 … 5 V and 0 … 10 V outputs | RL > 10 kΩ |
Relay outputs | |
Number and type of relays | 2 pcs, SPDT |
Max. switching power, current, voltage |
30 W, 1 A, 40 V DC / 28 V AC |
Max. wire size in PELV version | 2.5 mm2 (14 AWG) |
Max. wire size in AC (mains) version | 1.5 mm2 (16 AWG) |
Ethernet interface | |
Supported standards | 10BASE-T, 100BASE-TX |
Connector | 8P8C (RJ45) |
Supported protocols | Modbus TCP/IP (port 502), HTTPS (port 8443) |
Property | Description/Value |
---|---|
EU directives and regulations |
EMC Directive (2014/30/EU) Low Voltage Directive (2014/35/EU) RoHS Directive (2011/65/EU) amended by 2015/863 |
Electromagnetic compatibility (EMC) |
IEC/EN 61326‑1, industrial environment CISPR 32 / EN 55032, Class B |
Electrical safety |
IEC/EN 61010‑1 |
Type approvals |
DNV GL certificate no. TAA000032M |
Compliance marks |
CE, China RoHS, FCC, RCM, UKCA |
Listing marks |
UL Listed (USA and Canada) |
FCC compliance |
FCC Part 15, Class B |
Property | Description/Value |
---|---|
UL 50E (NEMA) rating |
NEMA 4: Hose down and splashing water, NEMA 250-2014 Type 4 |
Housing classification | IK08, DIN EN ISO 11997-1: Cycle B (VDA 621-415) |
Housing material | AlSi10Mg (DIN 1725) |
Display window material | Strengthened glass (IK08) |
Weight | 1.5 kg (3.3 lb) |
Dimensions (H × W × D) | 142 × 182 × 67 mm (5.63 × 7.17 × 2.64 in) |
Cable diameters for cable glands | |
M20×1.5 glands | 5.0 … 9.0 mm (0.20 … 0.31 in) |
M20×1.5 glands with split bushing | 7 mm (0.28 in) |
M16×1.5 glands | 2.0 … 6.0 mm (0.08 … 0.24 in) |
Property | Description/Value |
---|---|
User interfaces | Web interface for remote use, optional touchscreen display |
Supported languages | English, Chinese, French, German, Japanese, Spanish |
Optional display | 5" capacitive touchscreen |
Integrated data logging capabilities | Non-volatile memory, at least 10 years' storage with 24 h interval logging |
Measurement type | Probe models |
---|---|
Humidity and temperature | HMP1, HMP3, HMP4, HMP5, HMP7, HMP8, HMP9 |
Temperature | TMP1 |
Dew point | DMP5, DMP6, DMP7, DMP8 |
Carbon dioxide | GMP251, GMP252 |
Vaporized hydrogen peroxide | HPP271, HPP272 |
Moisture in oil | MMP8 |
Device or series | Models |
---|---|
MHT410 Moisture, Hydrogen and Temperature Transmitter | MHT410 |
Polaris™ Process Refractometers 6 | PR53AC, PR53AP, PR53GC, PR53GP, PR53M, PR53SD, PR53W |
Indigo80 Handheld Indicator | Indigo80 |