Indigo520 specifications - INDIGO520 - INDIGO500SERIES - INDIGO510

Indigo510, Indigo520 User Guide

Document code
M212287EN
Revision
M
Language
English (United States)
Product
INDIGO520
INDIGO500SERIES
INDIGO510
Document type
User guide
Table 1. Indigo520 transmitter options
Property Description/Value
Display
  • Touchscreen display
  • No display (indicator LED)
Powering
  • Protective extra-low voltage (15 … 35 V DC, 24 V AC ± 20%)
  • AC (mains) power (100 … 240 V AC 50/60 Hz)
  • Power over Ethernet (no analog outputs or relays)
Table 2. Indigo520 measurement performance
Property Description/Value
Barometric pressure (optional module)
Pressure range 500 … 1100 hPa
Class A:
Linearity ±0.05 hPa
Hysteresis ±0.03 hPa
Repeatability ±0.03 hPa
Calibration uncertainty ±0.07 hPa
Accuracy at +20 °C / +68 °F ±0.10 hPa
Temperature dependence ±0.1 hPa
Total accuracy (-40 … +60 °C / -40 … +140 °F) ±0.15 hPa
Long-term stability/year ±0.1 hPa
Response time (100 % response):
One sensor 2 s
Pressure units hPa, mbar, kPa, Pa, inHg, mmH20, mmHg, torr, psia
Table 3. Indigo520 operating environment
Property Description/Value
For use in wet locations Yes
Operating humidity

0 … 100 %RH

Maximum operating altitude, AC (mains) power

3000 m (9843 ft)

Maximum operating altitude, protective extra-low voltage (PELV) and Power over Ethernet (PoE) 4000 m (13 123 ft)

IP rating

IP66: Dust-tight. Protected from powerful water jets from any direction. 1

Operating temperature
With display −20 … +55 °C (−4 … +131 °F)
Without display −40 … +60 °C (−40 … +140 °F)
Without display with barometer module −40 … +55 °C (−40 … +131 °F)
Storage temperature
With display

−30 … +60 °C (−22 … +158 °F)

Without display

−40 … +60 °C (−40 … +140 °F)

Table 4. Indigo520 inputs and outputs
Property Description/Value
Operating power 2
Protective extra-low voltage (PELV) version 15 … 35 V DC, 24 V AC ±20 % 50/60 Hz, max. current 2 A (power supply input is galvanically isolated)

Fuse size for power supply: 3 A

Isolation voltage: 500 V AC, 1000 V DC

PELV power cable temp. rating ≥ +80 °C (+176 °F)
AC (mains) power version

100 … 240 V AC 50/60 Hz, max. current 1 A (power supply input is galvanically isolated)

Fuse size for power supply: 10 A

Isolation voltage: 1500 V AC

Power over Ethernet version 3

Power over Ethernet (PoE) IEEE 802.3at Type 2 Class 4

Max. current 600 mA, max. power consumption 25.5 W

Isolation voltage: 500 V AC, 1000 V DC

Analog outputs
Number of analog outputs 4
Selectable voltage output types 0 … 1 V, 0 … 5 V, 0 … 10 V, scalable
Selectable current output types 4 … 20 mA, 0 … 20 mA, scalable
Max. wire size 2.5 mm2 (14 AWG)
Accuracy of analog outputs at +20 °C (+68 °F) ±0.05 % full scale
Temperature dependence ±0.005 % / °C full scale
External loads:
Current outputs RL < 500 Ω
0 … 1 V output RL > 2 kΩ
0 … 5 V and 0 … 10 V outputs RL > 10 kΩ
Relay outputs
Number and type of relays 2 pcs, SPDT
Max. switching power, current, voltage

30 W, 1 A, 40 V DC / 28 V AC

Max. wire size in PELV version 2.5 mm2 (14 AWG)
Max. wire size in AC (mains) version 1.5 mm2 (16 AWG)
Ethernet interface
Supported standards 10BASE-T, 100BASE-TX
Connector 8P8C (RJ45)
Supported protocols Modbus TCP/IP (port 502), HTTPS (port 8443)
Table 5. Indigo520 compliance
Property Description/Value

EU directives and regulations

EMC Directive (2014/30/EU)

Low Voltage Directive (2014/35/EU)

RoHS Directive (2011/65/EU) amended by 2015/863

EMC compatibility

IEC/EN 61326‑1, industrial environment

CISPR 32 / EN 55032, Class B

Electrical safety

IEC/EN 61010‑1

Type approvals

DNV GL certificate no. TAA000032M

Compliance marks

CE, China RoHS, FCC, RCM

Listing marks

UL Listed (USA and Canada)

FCC compliance

FCC Part 15, Class B

Table 6. Indigo500 Series mechanical specifications
Property Description/Value

NEMA rating

NEMA 4: Hose down and splashing water, NEMA 250-2014 Type 4

Housing classification IK08, DIN EN ISO 11997-1: Cycle B (VDA 621-415)
Housing material AlSi10Mg (DIN 1725)
Display window material Strengthened glass (IK08)
Weight 1.5 kg (3.3 lb)
Dimensions (H × W × D) 142 × 182 × 67 mm (5.63 × 7.17 × 2.64 in)
Cable diameters for cable glands
M20×1.5 glands 5.0 … 9.0 mm (0.20 … 0.31 in)
M20×1.5 glands with split bushing 7 mm (0.28 in)
M16×1.5 glands 2.0 … 6.0 mm (0.08 … 0.24 in)
Table 7. Indigo500 Series user interfaces
Property Description/Value
User interfaces Web interface for remote use, optional touchscreen display
Supported languages English, Chinese, French, German, Japanese, Spanish
Optional display 5 in capacitive touchscreen
Integrated data logging capabilities Non-volatile memory, at least 10 years' storage with 24 h interval logging
1 Evaluated by Eurofins, not by UL.
2 The power supply option is selected when ordering the transmitter.
3 Transmitter shall be supplied by a Power Sourcing Equipment (PSE) unit which fulfils the requirements of IEEE802.3at specifications.