Product overview - CAB100

CAB100 Quick Guide

Document code
M212284EN
Revision
D
Language
English
Product
CAB100
Document type
User guide

Vaisala CMS Industrial Cabinet CAB100 is an instrument cabinet designed to integrate devices for measuring humidity, temperature, differential pressure, and other parameters.

CAB100 is available in 2 models, CAB100A (small enclosure) and CAB100B (large enclosure). The cabinets can be configured according to application requirements, with options for differential pressure transmitters, analog input channels for the connection of remote transmitters, and safety barriers or galvanic isolators for hazardous areas, used with intrinsically safe devices.

Table 1. CAB100 configuration options
CAB100 model Max. number of devices or measurement points per cabinet Communication interface
CAB100A
PDT101 model 4 PDT101 transmitters Serial port server or Vaisala vNet PoE data logger interface 1
Analog input channel model 4 analog input channels Serial port server
Safety barrier model 4 safety barriers

Galvanic isolator model

4 galvanic isolators
CAB100B 2
PDT101 model 16 PDT101 transmitters Serial port server
Analog input channel model 32 analog input channels
Safety barrier model 16 safety barriers
Galvanic isolator model 12 galvanic isolators
Table 2. CAB100 powering specifications
Property Description/Value
AC (mains) power

100 – 240 V AC, 50 – 60 Hz

0.5 A maximum (120 V AC)

Power supply module within cabinet 24 V DC / 2.5 A / Fused 2 A
Power over Ethernet 3

12 – 30 V DC

IEEE 802.3af, 10Base-T

Mains fuse (nominal) 4 A
DL4000 data loggers are delivered with default factory settings. This applies both to the factory-installed data loggers and the ones delivered as spare parts. For instructions on configuring measurement units and scaling in the data loggers, see the Vaisala viewLinc Enterprise Server User Guide for your viewLinc version, available at www.vaisala.com/viewlinc.
For the full CAB100 product specifications, installation instructions, wiring diagrams, and layout drawings, see www.vaisala.com/cab100.
1 vNet PoE interface in CAB100A PDT101 model only.
2 Mixed CAB100B models are also available, with certain limitations. Contact your Vaisala sales representative for details.
3 In CAB100A PDT101 model only.