| Window check (mode
windowcheck) does not change any values. |
|
Shortcut |
Description |
TAB |
Auto-completes a command, argument, parameter, or parameter value after
typing first characters, as far as typed input appears unique. |
TAB-TAB |
Prints available items. For example, type 's<TAB-TAB>' to see commands
beginning with 's'. |
|
You can cancel the process at any point.
-
Establish a maintenance connection to FD70.
For instructions, see FD70 Installation Guide
(M212221EN).
-
Type the following commands:
level 2
mode windowcheck
For the side receiver, type the
following command:
mode windowcheck 2
Follow the onscreen instructions.
| The window contamination values can
be slightly negative, which is normal. |
-
If prompted to clean the transmitter and receiver windows, do the following.
-
Using low pressure, spray 50 % isopropyl alcohol on the glass windows. Spray enough
solution to dissolve and loosen any dirt.
-
Wipe the window with a soft, lint-free cloth until the window is clean.
-
Take a new cloth and wipe the window until it is clean.
-
Wait until the windows are dry.
| During window cleaning, measurement
data is invalid. |
| During window cleaning, keep the
interface unit door closed. |
| CAUTION Do not use pressure
washers. |
| CAUTION Be careful not to
scratch the window surfaces or touch the measurement unit windows with bare
hands. |
The system monitors the window
contamination values during the cleaning.
Window contamination
values
Value |
Description |
-1.5 % > < 1.5 % |
Window check passed; windows are clean |
1.5–10 % |
Window cleaning is recommended |
> 10 % |
Window cleaning is mandatory |
You may need to clean the windows
twice.
-
If the cleaning is successful, no further actions are needed.
If the window check fails after
re-cleaning (the window contamination value stays above 1.5 %), the window transparency
may have permanently decreased.
-
A minor decrease in transparency can
be taken into account with calibration adjustment (mode adjust). See Performing calibration adjustment.
-
If the window transparency loss is
high or the window is damaged, a replacement hood is needed.
-
When done, close the maintenance connection.
The window check date is stored in the
system as metadata.