Available Parameters and Default Scaling - HMD60 - HMD65

HMD65 User Guide

Document code
M212243EN
Revision
C
Language
English
Product
HMD60
HMD65
Document type
User guide

HMD65 Measurement Parameters and Default Analog Output Scaling

Table 1 shows the available output parameters and the default analog output scaling of the parameters for HMD65.

Table 1. HMD65 Measurement Parameters and Default Scaling
Parameter Default Scaling for 1 … 10 V Output Range
Relative humidity (RH) 0 … 100 %RH
Temperature (T) -20 ... +80 °C (-4 ... +176 °F)
Dew point temperature (Td) -40 ... +80 °C (-40 ... +176 °F)
Dew point/frost point temperature (Tdf) -40 ... +80 °C (-40 ... +176 °F)
Absolute humidity (A) 0 … 300 g/m3 (0 … 131.1 gr/ft3)
Mixing ratio (X) 0 … 600 g/kg (0 … 4200 gr/lb)
Wet-bulb temperature (Tw) -40 ... +80 °C (-40 ... +176 °F)
Enthalpy (H) -40 … 1600 kJ/kg (-9.5 ... 695.6 Btu/lb)

Changing Measurement Parameter Scaling

If your application requires an analog output scaling that differs from the defaults shown in Table 1, you can configure the scaling by connecting the transmitter to Vaisala Insight PC software (requires Vaisala USB cable 219690).