CAUTION To preserve the IP rating of the enclosure, make sure that you lead cables through the flanges as instructed.
  • Wire cutters
  • Wire-stripping pliers
Figure 1. Stainless steel flange
1
Flange
2
Cable gland, cable diameter 7.5 … 14 mm (0.30 … 0.55 in)
3
Dummy plug M16 × 1.5
4
Cable gland, cable diameter 5 … 10 mm (0.20 … 0.39 in)
5
Cable gland, cable diameter 3 … 6.5 mm (0.12 … 0.26 in)
  1. Lead the cables through the cable glands in the flange.
  2. RJ45 connectors do not fit through the cable glands. Cut off the existing connector, lead the Ethernet cable through a cable gland, and attach a new cable connector to the cable. For instructions, see the connector documentation.
  3. Inside the enclosure, connect the cables and wires according to the wiring diagram.
CAUTION To avoid strain on the connector, make sure that you connect cables and wires properly. Too much strain may cause the cable or wire to fall off or damage the cable, wire, or connector.