Model | Measurement range | Output |
---|---|---|
PDT101-P4C | ±60 Pa | 4 … 20 mA |
PDT101-P4V | ±60 Pa | 0 … 5 V |
PDT101-P4C2 | ±125 Pa | 4 … 20 mA |
PDT101-P4V2 | ±125 Pa | 0 … 5 V |
PDT101-W4C | ±0.25 inH20 | 4 … 20 mA |
PDT101-W4V | ±0.25 inH20 | 0 … 5 V |
PDT101-W4C2 | ±0.5 inH20 | 4 … 20 mA |
PDT101-W4V2 | ±0.5 inH20 | 0 … 5 V |
Property | Description/Value |
---|---|
Measurement ranges (bidirectional) | ±60 Pa, ±125 Pa, ±0.25 inH2O, or ±0.5 inH2O |
Accuracy (incl. non-linearity, hysteresis, repeatability and zero/span calibration settings) | 0.4 % span |
Long-term stability | ≤ 0.5 % span/year |
Response time (10 ... 90 %) | 250 ms |
Warm-up time | 15 s |
Compensated temperature range |
+2 ... +54 °C (+35.6 ... +129.2 °F) |
Temperature dependence |
±(0.065 Pa + 0.054 % of reading) / °C or ±(0.00015 inH2O + 0.03 % of reading) / °F (reference 21 °C or 70 °F) |
Pressure type |
Differential, gauge, vacuum and compound
|
Overpressure | |
Proof pressure Burst pressure Static pressure |
1.0 bar 1.7 bar 1.7 bar |
Mounting position | |
Error (zero adjustable) | ≤ 1 %/g (calibration in vertical position is standard) |
Adjustments (front accessible) | |
Zero Span |
±5 % span ±3 % span |
Property | Description/Value |
---|---|
Operating temperature | −18 ... +70 °C (−0.4 ... +158 °F) |
Storage temperature | −40 ... +82 °C (−40 ... +179.6 °F) |
Note: If used in an electromagnetic field of 3 V/m, with narrow frequency area of 80 ... 120 Mhz, it is possible that the current output of PDT101 can deviate max. 0.8 % (with accuracy specified 0.4 %) |
Property | Value/Description |
---|---|
EMC compatibility |
EN 61326‑1, basic immunity test requirements |
Compliance marks |
CE, RCM |
Property | Description/Value |
---|---|
Process connection | 1/4" barbed fittings |
Max. loop resistance for 4 ... 20 mA | ≤ (Supply voltage - 12 V)/0.022 A |
Supply current | Max. 20 mA for 4 ... 20 mA output signal |
Optical process diagnostics | LED visual indicator |
Electrical connection |
Euro style pluggable terminal block accepts 12 ... 26 AWG wire (0.13 up to 3.31 mm2) |
Output signal | |
2-wire 3-wire |
4 ... 20 mA 0 ... 5 V DC (user selectable 0 ... 10 V DC) |
Operating voltage | |
2-wire output 4 ... 20 mA 3-wire output 0 ... 5 V DC 3-wire output 0 ... 10 V DC |
12 ... 36 V DC 11.5 ... 36 V DC or 24 V AC 14 ... 36 V DC or 24 V AC |
Property | Description/Value |
---|---|
Medium (measured gas) | Clean and dry air, non-conducting and non-corrosive gases |
Mounting | Threaded fastener for wall mounting or DIN rail type EN 50022 |
IP rating | IP40 |
Weight | 0.07 kg |
Material | |
Process connection Sensor element Case |
Brass Silicon, aluminum, glass NEMA type 1 fire-retardant ABS 1 (meets UL94-5VA) |