Replacing cables of external sensors - FD70 - HMP155 - LM21

FD70 Configuration and Maintenance Guide

Document code
M212222EN
Revision
J
Language
English
Product
FD70
Document type
User guide
  • Screwdriver
  • Spanner 33/22 mm
  • Side-cutting pliers (if cable ties in use)

Make sure that you have the replacement spare part available.

Follow this instruction when replacing the cables of HMP155 and LM21.
  1. Switch off both AC (mains) power switch and battery switch.

    Make sure that there is no AC (mains) power available for the system.

    1
    Battery switch
    2
    AC (mains) power switch
  2. Disconnect the cable from the connector on the sensor.
  3. If the cable is attached with cable ties to the mast and/or sensor support arm, remove the cable ties.
  4. In the interface unit, loosen the cable from the cable gland.
  5. Disconnect the wires from the connector and remove the cable.
    1
    HMP155 wiring
    2
    LM21 wiring
  6. Route the replacement cable through the cable gland.
  7. Connect the wires of the replacement cable in the connectors.
    The board includes markings to indicate how to connect the wires.
    HMP155 wiring
    Wire Color Description
    Red GND
    Blue Vin+
    VH+(F1)
    VH−(F1)
    Brown COM3A−
    Pink COM3B+
    LM21 wiring
    Wire color Description
    Gray GND
    Pink Vin+
    White VH+(F2)
    Brown VH−(F2)
    Yellow COM3A−
    Green COM3B+
    D/N
    GND
  8. Route the cable from the interface unit to the sensor, attaching the cable with cable ties to the mast and/or sensor support arm.
  9. Power up the system.
  10. Verify operation.