FS11 polling - FD70

FD70 Configuration and Maintenance Guide

Document code
M212222EN
Revision
J
Language
English
Product
FD70
Document type
User guide

The FS11 poll command contains a maximum of 8 characters depending on the options used. See the examples of FS11 polling below.

EQFSiNNCR
FS11 poll command
Position Name Description Value/Range
1 EQ Enquiry 0x05
2 - 3 FS Prefix Fixed string = "FS" or "fs"
4 i Unit identifier 0 … 9, a … z, A … Z
5-6 NN Message number 00 … 99
7 CR Carriage return 0x0D

FS11 polling examples

The device has no unit identifier set. A default message is polled:

EQFS CR

The device has no unit identifier set. A message number 5 is polled.

EQfs 05CR

The device has unit identifier set = 2. A message set as default is polled:

EQFS2CR

The device has unit identifier set = 3. A message number 5 is polled.

EQFS305CR

FS11 data message example

SHFS3SXVIS(02000(AL(0)))BL(00100(AL(0)))<ETX>CSUMETCRLF