Replacing surge protection unit QSP431 - FD70 - QSP431

FD70 Configuration and Maintenance Guide

Document code
M212222EN
Revision
J
Language
English
Product
FD70
Document type
User guide

Make sure that you have the replacement spare part available.

If the connectors of the existing component are still usable, do not disconnect the wiring, but remove the connectors and use them for the replacement component.
Replacing surge protector unit QSP431
1
Battery switch
2
AC (mains) power switch
3
Surge protector (for RS-232/RS-485 data)
  1. Switch off both AC (mains) power switch and battery switch.

    Make sure that there is no AC (mains) power available for the system.

  2. Disconnect the wiring and pull out the component from the DIN rail.
  3. Connect the wires.
    Serial data cable wiring: RS-232
    QSP431 Pin Data cable
    TxD 4 TXD (data out)
    GND 5 Signal ground
    RxD 6 RXD (data in)
    Serial data cable wiring: RS-485
    QSP431 Pin Data cable
    D+ 4 RS-485 D+
    GND 5 Signal ground
    D− 6 RS-485 D−
    QSP431 internal wiring
    X2 Pin Color Description
    4 Green TXD/D+
    5 Dark blue Signal ground
    6 Yellow RXD/D−
  4. Attach the replacement component on the DIN rail.
  5. Attach the connectors on the connector board.

    The components for data 1 and data 2 are side by side on the DIN rail. Data 1 is on the left and data 2 on its right.

  6. Power up the system.