FD70 has a transmitter, a main receiver and a side receiver, depending on the configuration
you have chosen. The following procedure describes the general calibration checking steps.
Perform calibration for the main receiver and transmitter, and then perform the same
calibration steps for side receiver and transmitter.
Before you start, clean the windows as instructed in section Cleaning windows.
During window cleaning, measurement
data is invalid.
Calibration check (mode
calcheck, mode calcheck 2) does not change any values.
Requirements for calibration
check
Property
Requirement
Visibility
> 2000 m (6560 ft)
Present weather
Calibration is not recommended to be carried out in heavy rain or
in bright sunshine
Perform calibration check:
At least every 12 months.
In case of severe physical or mechanical
impact.
Calibration check has the following
phases:
Preparation
Zero signal calibration check
Scatter signal calibration check
The result of the calibration check is one of the following:
Pass: the calibration is valid
Fail: calibration adjustment is needed
You can cancel the process at any point.
Before you start, check the opaque glass of the calibrators and clean them if
necessary.
Be careful not to scratch the
calibrators, especially the opaque glass. Damaged calibrators have a negative effect on
the calibration.
Establish a maintenance connection to FD70.
Type the following commands:
level 2
mode calcheck
For the side receiver, type the
following command:
mode calcheck 2
Follow the onscreen instructions.
When prompted for Calibrator Signal Value, type the 4‑digit value.
The same value is printed on the
calibrator certificate, on the calibrator labels, and the calibrator case.
When prompted to install the calibrators with zero signal plates, remove the protection
caps from the transmitter and receivers.
Attach the calibrators with zero signal plates on the transmitter and receiver.
There is a dedicated calibrator for the
transmitter and one for the receivers. Check the labels on the calibrators to make sure
that you attach them correctly. The following figures show the calibrator attachment
process for the main receiver and the side receiver.
The system performs zero signal
calibration.
The zero signal calibration passes
if the zero signal value is within ±0.012.
If the zero signal calibration
fails, eliminate all possible reflections and external light sources.
When prompted, remove the zero signal plates.
The system performs scatter signal
calibration. Keep the measurement area clear to avoid interference.
The scatter signal calibration
passes if the scatter signal value is within 3 % of the calibrator signal value.
If the scatter signal calibration
fails, calibration adjustment is needed (mode adjust,
mode adjust 2).
When done, remove the calibrators, reattach the protection caps on the transmitter and
receiver, and close the maintenance connection.