Status message - FD70

FD70 Configuration and Maintenance Guide

Document code
M212222EN
Revision
J
Language
English
Product
FD70
Document type
User guide

You can poll a human-readable status message from FD70 using the command STATUS through the CLI. The status message contains information about:

  • Device configuration
  • Maintenance alerts
  • Hardware failure alerts
The character numbers (flags) not listed in the tables below are either reserved for future use or are defined but not implemented. These flags have a fixed value (0).
Block 1 status message fields: device configuration
Alert Position (status word character number) Value Description
Recent re-start occurred 3 0 Normal operation
I Device has recently been re-started

External RH+T Sensor HMP155

External source HMP155 is activated by parameter com3.hmp155.data

4 0 Activated
I Not activated

Ambient temperature source

Indicates the source of the ambient temperature observations reported

5 0 Internal
I External

Background Luminance Sensor LM21

6 0 Activated
I Not activated
Day / Night switch input 7 0 Activated
I Not activated
Contamination compensation 8 Reserved for future use.
Flying insect compensation 9 0 Activated (default)
I Disabled

Heating control mode

10 0

Automatic

Heaters are activated or inactivated automatically based on temperature.

I Forced active
W Forced inactive

Heating power mode

11 0 Normal
I Low
Block 2 status message fields: maintenance
Alert Position (status word character number) Value Description Actions

Maintenance overall

  • Overall maintenance alert is raised according to the highest severity in this block.

15 0 OK
I At least 1 maintenance indication Prepare for maintenance. Check the details. Actions may be needed.
W At least 1 maintenance warning Maintenance recommended. Check the details. Actions recommended.
A At least 1 maintenance alarm Immediate maintenance needed. Check the details. Immediate action needed.
Calibration 16 0 OK
A Calibration ongoing or invalid
  • Calibration ongoing: Do not switch off the unit.
  • Calibration invalid: Recalibrate. If recalibration fails, contact Vaisala Service Center.
Clean window reference adjustment 17 0 OK
W Recommended
A Needed See Performing calibration adjustment.
Contamination - main path 18 0 OK
  • Slight: Prepare for window cleaning in near future. Measurement data is valid.
  • Moderate: Cleaning is recommended. Measurement data is valid.
  • Dirty: Immediate windows cleaning needed. Measurement data is NOT valid.
I Slight
W Moderate
A Dirty
Contamination - side path 19 0 OK
I Slight
W Moderate
A Dirty
Contamination - transmitter 20 0 OK
I Slight
W Moderate
A Dirty
Contamination - main receiver 21 0 OK
I Slight
W Moderate
A Dirty
Contamination - side receiver 22 0 OK
I Slight
W Moderate
A Dirty
Blocking - transmitter 23 0 OK
W Intermittently blocked
A Continuously blocked
Blocking - main receiver 24 0 OK
W Intermittently blocked
A Continuously blocked
Blocking - side receiver 25 0 OK
W Intermittently blocked
A Continuously blocked
Ambient light - main receiver 27 0 OK
W Interference

Interference or saturation caused by excessive level of steady ambient light.

Eliminate the interfering light source or adjust the orientation of the measurement unit, for example, to avoid direct or indirect sunlight.

A Saturation
Ambient light - side receiver 28 0 OK
W Interference
A Saturation
Operating voltage 31 0 OK
W Low/High Operating voltage outside normal range. Check the condition of the power supply.
Heating voltage 32 0 OK
W Low/High Heater voltage outside normal range. Check the condition of the power supply.
Power supply 33 0 Normal
I Backup battery Device is running on battery backup with no heating. Be aware of remaining operating time and prepare for needed actions.
Calibration - background luminance sensor 35 0 OK
  • Calibration ongoing: LM21 is being calibrated. The alarm will disappear when the LMA21 calibrator is disconnected. The observations of FD70 during the calibration may still be valid.
  • Calibration failure: Calibration parameters have been lost. Re-calibrate using LMA21, calibrate the window contamination measurement and check parameters.
I Calibration ongoing
W Calibration failure
Contamination - background luminance sensor 36 0 OK
  • Dirty: Window cleaning needed.
W Dirty
Blocking - background luminance sensor 37 0 OK
  • Blocking: Appearing object blocking the light path. Remove any obstacles in front of the window and clean the window.
W Blocking
Block 3 status message fields: hardware
Name Position (status word character number) Value Description Actions

Hardware overall

  • Overall hardware alert is raised according to the highest severity (warning or alarm) on any hardware sub-unit in this block, for example transmitter. 1

41 0 OK
W At least one warning on some sub-unit Actions recommended.
A At least one alarm on some sub-unit Immediate action needed.

Transmitter

  • Overall sub-unit alert is raised according to the highest severity alert (warning or alarm) of the sub-unit. 1

42 0 OK
W Failure Check the alert details.
A Failure

Main receiver

  • Overall sub-unit alert is raised according to the highest severity alert (warning or alarm) of the sub-unit. 1

43 0 OK
W Failure Check the alert details.
A Failure

Side receiver

  • Overall sub-unit alert is raised according to the highest severity alert (warning or alarm) of the sub-unit. 1

44 0 OK
W Failure
A Failure
Controller
  • Overall sub-unit alert is raised according to the highest severity alert (warning or alarm) of the sub-unit. 1

45 0 OK
W Failure Check the alert details.
A Failure Check the alert details.
External sensor HMP155
  • Overall sub-unit alert is raised according to the highest severity alert (warning or alarm) of the sub-unit. 1

46 0 OK
W Failure Check the alert details.
External sensor LM21
  • Overall sub-unit alert is raised according to the highest severity alert (warning or alarm) of the sub-unit. 1

47 0 OK
W Failure Check the alert details.
Block 4 status message fields: transmitter
Name Position (status word character number) Value Description Actions
Transmitter hood heater 50 0 OK, heater inactive
I OK, heater active
W Failure
Transmitter hood temperature sensor 51 0 OK
W Failure
Transmitter window lifespan 52 0 OK
W Decreased
A Expired
Transmitter light source age 53 0 OK
W Warning
Transmitter data communication 54 0 OK
A Failure
TX optical power 55 0 OK
A May exceed safety limits
Transmitter module electronics 56 0 OK
W Failure Replace the module. See Replacing transmitter module.
A Failure
57 + Delimiter
Transmitter electronics failure code 58–60 000 ‑ FFF Reserved for future use.
Block 5 status message fields: main receiver
Name Position (status word character number) Value Description Actions
Main receiver hood heater 62 0 OK, heater inactive
I OK, heater active
W Failure
Main receiver hood temperature sensor 63 0 OK
W Failure
Main receiver window lifespan 64 0 OK
W Decreased
A Expired
Main receiver signal communication 65 0 OK
A Failure
Main receiver data communication 66 0 OK
A Failure
Main receiver module electronics 68 0 OK
W Failure Replace the module. See Replacing main receiver module.
A Failure
69 + Delimiter
Main receiver electronics failure code 70–72 000 ‑ FFF
The following block is not displayed in the status command if the FD70 model does not include side receiver.
Block 6 status message fields: side receiver
Name Position (status word character number) Value Description Actions
Side receiver hood heater 74 0 OK, Heater Inactive
I OK, Heater Active
W Failure
Side receiver hood temperature sensor 75 0 OK
W Failure
Side receiver window lifespan 76 0 OK
W Decreased
A Expired
Side receiver signal communication 77 0 OK
A Failure
Side receiver data communication 78 0 OK
A Failure
Dry obscuration 79 0 Not detected
I Detected
Side receiver module electronics 80 0 OK
W Failure Replace the module. See Replacing main receiver module.
A Failure
81 + Delimiter
Side receiver electronics failure code 82–84 000 ‑ FFF
Block 7 status message fields: controller
Name Position (status word character number) Value/Range Description Actions
Ambient temperature sensor failure 86 0 OK
A Temperature sensor failure Replace the internal ambient temperature sensor.
Algorithm alert 87 0 OK Set a longer autosend interval.
W Measurement data missing
Controller module electronics 89 0 OK
A Failure Contact Vaisala.
90 + Delimiter
Controller electronics failure code 91–95 00000 ‑ FFFFF
The following block is not displayed in the status command if HMP155 is not activated.
Block 8 status message fields: external sensor HMP155
Name Position (status word character number) Value/Range Description Actions
HMP155 temperature measurement 97 0 OK
W Failure Replace HMP155.
HMP155 temperature probe measurement 98 0 OK
W Failure
HMP155 humidity measurement 99 0 OK
W Failure Check HUMICAP sensor.
HMP155 data communication 101 0 OK
W Failure Check sensor connection.
HMP155 electronics 103 0 OK
W Failure Replace HMP155.
104 + Delimiter
HMP155 electronics failure code 105–107 000 ‑ FFF
The following block is not displayed in the status command if LM21 is not activated.
Block 9 status message fields: LM21
Name Position (status word character number) Value/Range Description Actions
LM21 hood heater 109 0 OK
W Failure Replace the internal ambient temperature sensor.
LM21 data communication 113 0 OK
W Failure Contact Vaisala.
LM21 electronics 115 0 OK
W
+ Delimiter
LM21 failure code 117–119

Status message example

0> status

Device configuration
  Recent re-start occurred        :0: Normal operation
  External sensor HMP155          :0: Activated
  Ambient temperature source      :I: External
  External sensor LM21            :0: Activated
  Day / night switch input        :I: Disabled
  Flying insect compensation      :0: Enabled
  Heating control mode            :0: Automatic
  Heating power mode              :0: Normal

Maintenance
  Maintenance overall             :0: OK
  Calibration                     :0: OK 2022-02-07
  Clean window reference          :0: OK
  Contamination - path            :0: OK 5 %
  Contamination - side path       :0: OK -1 %
  Contamination - TX              :0: OK -1 %
  Contamination - RX              :0: OK 5 %
  Contamination - side RX         :0: OK 0 %
  Blocking - TX                   :0: OK
  Blocking - RX                   :0: OK
  Blocking - side RX              :0: OK
  Ambient light - RX              :0: OK
  Ambient light - side RX         :0: OK
  Operating voltage               :0: OK
  Heating voltage                 :0: OK
  Power supply                    :0: Normal
  LM21 calibration                :0: OK
  LM21 contamination              :0: OK
  LM21 blocking                   :0: OK

Hardware
  Hardware overall                :0: OK
  Transmitter                     :0: OK
  Receiver                        :0: OK
  Side receiver                   :0: OK
  Controller                      :0: OK
  External sensor HMP155          :0: OK
  External sensor LM21            :0: OK

Transmitter
  TX hood heater                  :0: OK
  TX hood temperature sensor      :0: OK
  TX window lifespan              :0: OK
  TX light source age             :0: OK
  TX data communication           :0: OK
  TX optical power                :0: OK
  TX module electronics           :0: OK Failure Code=000

Receiver
  RX hood heater                  :I: Active
  RX hood temperature sensor      :0: OK
  RX window lifespan              :0: OK
  RX data communication           :0: OK
  RX module electronics           :0: OK Failure Code=000

Side Receiver
  Side RX hood heater             :0: OK
  Side RX hood temperature sensor :0: OK
  Side RX window lifespan         :0: OK
  Side RX data communication      :0: OK
  Dry obscuration                 :0: Not detected
  Side RX module electronics      :0: OK Failure Code=000

Controller
  Internal ambient T sensor       :0: OK
  Algorithm                       :0: OK
  Controller module electronics   :0: OK Failure Code=00000

External sensor HMP155
  HMP155 temperature measurement  :0: OK
  HMP155 humidity measurement     :0: OK
  HMP155 data communication       :0: OK
  HMP155 electronics              :0: OK Failure Code=000

External sensor LM21
  LM21 hood heater                :0: OK
  LM21 data communication         :0: OK
  LM21 electronics                :0: OK Failure Code=000

0>
1 Blocks from 4 onwards contain the detailed alerts of each sub-unit.