Make sure that you have the replacement spare part available.
-
Switch off both AC (mains) power switch and battery switch.
Make sure that there is no AC (mains) power available for the system.
-
Open the 2 screws but do not remove them. Pull them approximately 5 mm out from the
mast or sensor support arm.
- Lift the measurement unit carefully up, and cut the cable tie that holds the rubber cover in place. Remove the rubber cover.
- Disconnect the cable from the measurement unit.
- Connect the cable to the replacement measurement unit. Attach the rubber cover with a cable tie.
-
Mount the replacement measurement unit in place and fix with the 2 screws.
For further information, see FD70 Installation Guide (M212221EN).
- Configure the settings using terminal software. See FD70 Configuration and Maintenance Guide (M212222EN).
- Power up the system.
- Perform calibration adjustment (mode adjust) with the field calibrator kit (FDA701).