The following parameters are used with the GET, GETSET, and SET commands.
The following levels are in use:
- Level 0: read access
- Level 1: write access to basic parameters
- Level 2: write access to advanced parameters
To change the user level, use the level command.
Columns GET and SET indicate the required user level to view or change the setting.
Parameter | GET | SET | Description | Value / Range | Default |
---|---|---|---|---|---|
com1.func.type | 0 | 1 | Type of port |
|
RS485 |
com2.func.type | |||||
com1.func.mode | 0 | 1 | Mode of port |
|
DATA |
com2.func.mode | |||||
com1.func.delay | 0 | 1 |
Delay before transmit Unit: milliseconds |
0–1000 | 20 |
com2.func.delay | |||||
com1.func.termination | 0 | 1 | RS-485 bus termination resistor control ON/OFF |
|
ON |
com2.func.termination | |||||
com1.cli.timeout | 0 | 1 |
Timeout for CLI return to DATA mode Unit: minutes |
0–30 | 10 |
com2.cli.timeout | |||||
com1.cli.echo | 0 | 1 | CLI ECHO |
|
OFF |
com2.cli.echo |
Parameter | GET | SET | Description | Value / Range | Default |
---|---|---|---|---|---|
messages.mes8.line4enable | 0 | 1 | Add size distribution line to the end of data message |
|
ON |
messages.mes8.line5enable | 0 | 1 | Add fall speed distribution line to the end of data message |
|
ON |
Parameter | GET | SET | Description | Value / Range | Default |
---|---|---|---|---|---|
com1.comm.baud | 0 | 1 | Data speed |
|
9600 |
com2.comm.baud | |||||
com1.comm.data | 0 | 1 | Number of data bits |
|
8 |
com2.comm.data | |||||
com1.comm.parity | 0 | 1 | Parity |
|
N |
com2.comm.parity | |||||
com1.comm.stop | 0 | 1 | Number of stop bits |
|
1 |
com2.comm.stop |
Parameter | GET | SET | Description | Value / Range | Default |
---|---|---|---|---|---|
com1.message1.type | 0 | 1 | Default data message |
|
7 |
com2.message1.type | |||||
com1.message1.emulation | 0 | 1 | Message emulation of default message |
|
FD12 |
com2.message1.emulation | |||||
com1.message1.framing | 0 | 1 | Data message framing of default message |
|
ON |
com2.message1.framing | |||||
com1.message1.interval | 0 | 1 |
Autosend message interval Unit: seconds |
0 ... 3600 | 15 |
com2.message1.interval | |||||
com1.message1.validonalarm | 0 | 1 | Show observation data even when an ALARM is on |
|
OFF |
com2.message1.validonalarm | |||||
com1.message1.status | 0 | 1 | Add status line at the end of message. FS11 = FS11-compatible status line. |
|
OFF |
com2.message1.status |
Parameter | GET | SET | Description | Value / Range | Default |
---|---|---|---|---|---|
com3.func.delay | 0 | 2 |
Delay before transmit Unit: milliseconds |
0–1000 | 20 |
com3.comm.baud | 0 | 2 | Data speed |
|
9600 |
com3.comm.data | 0 | 2 | Number of data bits |
|
8 |
com3.comm.parity | 0 | 2 | Parity |
|
N |
com3.comm.stop | 0 | 2 | Number of stop bits |
|
1 |
com3.lm21.data | 0 | 1 | LM21 activate |
|
OFF |
com3.hmp155.data | 0 | 1 | HMP155 activate |
|
OFF |
Parameter | GET | SET | Description | Value / Range | Default |
---|---|---|---|---|---|
eth0.address.mac | 0 | − | MAC address | − | Set at factory, different for each device |
eth0.address.dhcp | 0 | 1 | DCHP protocol |
|
OFF |
eth0.address.ip1 | 0 | 1 | IP address | − | 192.168.2.101 |
eth0.address.nm1 | 0 | 1 | Subnet mask | − | 255.255.255.0 |
eth0.address.gw1 | 0 | 1 | Internet gateway | − | 192.168.2.1 |
eth0.address.dns1 | 0 | 1 | DNS server 1 | − | 192.168.2.1 |
eth0.address.dns2 | 0 | 1 | DNS server 2 | − | − |
eth0.cli.timeout | 0 | 1 | Timeout in minutes for CLI return to DATA mode. Zero to disable. | 0–30 | |
eth0.message1.type | 0 | 1 | Data message send in autosend mode, or when polled for default message |
|
|
eth0.message1.emulation | 0 | 1 | Message emulation of default message |
|
|
eth0.message1.framing | 0 | 1 | Add message framing to data messages |
|
|
eth0.message1.interval | 0 | 1 | Message send interval in seconds when in autosend mode. 0 for poll only mode | 0–3600 | |
eth0.message1.validonalarm | 0 | 1 | Show observation fata even when an ALARM is on (= do not "slash" the data) |
|
|
eth0.message1.status | 0 | 1 | Add status line to the end of data message. FS11 = FS11-compatible status line. |
|
Parameter | GET | SET | Description | Value / Range | Default |
---|---|---|---|---|---|
time.sync.ntp1 | 0 | 1 | Network Time Protocol server address 1 |
|
2.openembedded.pool.ntp.org |
time.sync.ntp2 | 0 | 1 | Network Time Protocol server address 2 | OFF | |
time.unit.optime | 0 | − | Unit operation hours since factory reset |
Format: <days>d <hours>h |
|
time.unit.now | 0 | 1 | Date and time | Format: YYYY-MM-DD HH:MM:SS UTC |
− |
Parameter | GET | SET | Description | Value / Range | Default |
---|---|---|---|---|---|
io1.daynightinput.data | 0 | 1 | Day/night switch activate |
|
OFF |
Parameter | GET | SET | Description | Value / Range | Default |
---|---|---|---|---|---|
heater.hood.mode | 0 | 1 | Hood heater mode |
|
AUTO |
heater.hood.powerMode | 0 | 1 | Hood heater power mode |
|
NORMAL |
heater.dew.mode | 0 | 1 | Dew heater mode |
|
ON |
Parameter | GET | SET | Description | Value / Range | Default |
---|---|---|---|---|---|
heater.hood.Tmax | 2 | 2 |
When air temperature is above this, it is used as input for AUTO mode Unit: Celsius |
-60 ... 60 | 40 |
heater.hood.Tmin | 2 | 2 |
When air temperature is below this, it used as input for AUTO mode Unit: Celsius |
-60 ... 60 | 10 |
heater.hood.dTon | 2 | 2 |
In AUTO mode, when hood temperature is below air temperature plus dTon offset, individual hood heating is switched on Unit: Celsius |
-60 ... 60 | 5 |
heater.hood.dToff | 2 | 2 |
In AUTO mode, when hood temperature is above air temperature plus dToff offset, individual hood heating is switched off Unit: Celsius |
-60 ... 60 | 7 |
heater.hood.Tsafe | 2 | 2 |
Hood heater safety limit Maximum temperature of individual hoods at which hood heating and dew heating is applied in any conditions Unit: Celsius |
-60 ... 65 | 50 |
heater.hood.cycletime | 2 | 2 |
Hood heater cycling time. Unit: seconds |
5–20000 | 30 |
heater.hood.tempcompensation | 2 | 2 | Compensation time added to heating/pause cycle time for temperature difference to other hoods. Hood is heated/paused for temperature compensation seconds less or longer per 1 degree Celsius difference to mean of other hoods. | Seconds/degree Celsius | 5.0 |
heater.hood.cyclemin | 2 | 2 |
Minimum cycle time under which temperature compensation does not adjust cycle time. Unit: seconds. |
[5, 20000] | 15 |
heater.dew.Ton | 2 | 2 |
In dew heater ON mode, when air temperature is below this, dew heater is switched on Unit: Celsius |
-60 ... 60 | 45 |
heater.dew.Toff | 2 | 2 |
In dew heater ON mode, when air temperature is above this, dew heater is switched off Unit: Celsius |
-60 ... 60 | 47 |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
fd.product.itemcode | 0 | − | Product item code | FD70 |
fd.product.fullname | 0 | − | Product full name | Forward Scatter Sensor FD70 |
fd.product.model | 0 | − | Product model | |
fd.product.modelname | 0 | − | Product model full name |
|
fd.product.revision | 0 | − | Product revision | A–Z |
fd.product.pw | 0 | − | Present weather support |
|
fd.product.mor | 0 | − |
Visibility range Unit: kilometers |
|
fd.product.disdro | 0 | − |
|
Parameter | GET | SET | Description | Value / Range | Default |
---|---|---|---|---|---|
fd.unit.deviceid | 0 | 1 |
Device identifier In MES 7 and MES 8, known as unit identifier. |
|
− |
fd.unit.alias1 | 0 | 1 | Alias name 1, max. 80 characters | Freeform text | − |
fd.unit.alias2 | 0 | 1 | Alias name 2, max. 80 characters | Freeform text | − |
fd.unit.latitude | 0 | 1 |
Latitude Unit: degrees |
−90.000000 … 90.000000 | 0.0 |
fd.unit.longitude | 0 | 1 |
Longitude Unit: degrees |
−180.000000 ... 180.000000 | 0.0 |
fd.unit.altitude | 0 | 1 |
Altitude Unit: meters Resolution: full meters Use negative value (-) for values below sea level |
−999 ... 9999 | 0 |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
fd.unit.serial | 0 | − | Serial number | − |
fd.unit.configuration | 0 | − | Configuration code | − |
fd.unit.mfgdate | 0 | − | Manufacturing date | − |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
fsc.product.itemcode | 0 | − | Controller module item code | FSC301 |
fsc.product.revision | 0 | − | Controller module revision | A–Z |
fsc.unit.serial | 0 | − | Controller module serial number | − |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
fsrm.product.itemcode | 0 | − | Main receiver module item code | FSR301 |
fsrm.product.revision | 0 | − | Main receiver module revision | A–Z |
fsrm.unit.serial | 0 | − | Main receiver module serial number | − |
fsrm.unit.mfgdate | 0 | − | Main receiver module manufacturing date | − |
fsrm.factory.calibrationdate | 0 | − | Date of factory calibration of main receiver (clean reference set) | − |
fsrs.product.itemcode | 0 | − | Side receiver module item code | |
fsrs.product.revision | 0 | − | Side receiver module revision | |
fsrs.unit.serial | 0 | − | Side receiver module serial number | |
fsrs.unit.mfgdate | 0 | − | Side receiver module manufacturing date | |
fsrs.factory.calibrationdate | 0 | − | Factory calibration date |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
fss.product.itemcode | 0 | − | Surge board module item code | |
fss.product.revision | 0 | − | Surge board module revision | |
fss.unit.serial | 0 | − | Surge board module serial number | |
fss.unit.mfgdate | 0 | − | Surge board module manufacturing date |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
fst.product.itemcode | 0 | − | Transmitter module item code | FST301 |
fst.product.revision | 0 | − | Transmitter module revision | A–Z |
fst.unit.serial | 0 | − | Transmitter module serial number | − |
fst.unit.mfgdate | 0 | − | Transmitter module manufacturing date | − |
fst.factory.calibrationdate | 0 | − | Date of factory calibration of transmitter | − |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
lm.product.code | 0 | − | Background luminance sensor product model | LM21 |
lm.product.serial | 0 | − | Background luminance sensor serial number | − |
lm.product.sw | 0 | − | Background luminance sensor software version | − |
hmp.product.code | 0 | − | Humidity and temperature sensor product model | HMP155 |
hmp.product.serial | 0 | − | Humidity and temperature sensor serial number | − |
hmp.product.sw | 0 | − | Humidity and temperature sensor software version | − |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
mor.datavalid.1min | 2 | 2 | 1 min visibility data validity | 1–100 |
mor.datavalid.10min | 2 | 2 | 10 min visibility data validity | 1–100 |
precip.datavalid.intensity | 2 | 2 | Precipitation intensity data validity | 1–100 |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
precipamount.scaling.liquid | 2 | 2 | Liquid scaling | |
precipamount.scaling.solid | 2 | 2 | Solid scaling | |
precipamount.conversion.snow | 2 | 2 | Snow conversion | |
precipamount.autozero.lwc | 2 | 2 | Precipitation accumulation zero | 0–999.99 |
precipamount.autozero.snow | 2 | 2 | Snow accumulation zero | 0–9999 |
precipamount.current.lwc | 2 | 2 | Current precipitation accumulation | 0–999.99 |
precipamount.current.snow | 2 | 2 | Current snow accumulation | 0–9999 |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
preciptype.snow.usevisibility | 2 | 2 | Use visibility |
|
preciptype.drizzle.moderate | 2 | 2 | Moderate drizzle limit | |
preciptype.drizzle.heavy | 2 | 2 | Heavy drizzle limit | |
preciptype.freezingdrizzle.moderate | 2 | 2 | Moderate freezing drizzle limit | |
preciptype.freezingdrizzle.heavy | 2 | 2 | Heavy drizzle limit | |
preciptype.rain.moderate | 2 | 2 | Moderate rain limit | |
preciptype.rain.heavy | 2 | 2 | Heavy rain limit | |
preciptype.rain.violent | 2 | 2 | Violent rain limit | |
preciptype.freezingrain.moderate | 2 | 2 | Moderate freezing rain limit | |
preciptype.freezingrain.tmin | 2 | 2 | Minimum freezing rain temperature | [-99.0 … 0.0] |
preciptype.freezingrain.heavy | 2 | 2 | Heavy freezing rain limit | |
preciptype.drizzlerain.moderate | 2 | 2 | Moderate drizzle and rain limit | |
preciptype.rainsnow.moderate | 2 | 2 | Moderate rain and snow limit | |
preciptype.snow.moderate | 2 | 2 | Moderate snow limit | |
preciptype.snow.heavy | 2 | 2 | Heavy snow limit | |
preciptype.snowvis.moderate | 2 | 2 | Moderate snow visibility limit | 0–100000 |
preciptype.snowvis.heavy | 2 | 2 | Heavy snow visibility limit | 0–100000 |
preciptype.icepellets.moderate | 2 | 2 | Moderate ice pellets limit | |
preciptype.icepellets.heavy | 2 | 2 | Heavy ice pellets limit | |
preciptype.limit.detection | 2 | 2 | Detection limit | |
preciptype.limit.mistvisibility | 2 | 2 | Mist visibility limit | 0–100000 |
preciptype.limit.mistvisibilitymetar | 2 | 2 | Mist visibility limit for METAR | 0–100000 |
preciptype.metar.fzrasn | 2 | 2 | Type reported in METAR when FZRASN detected |
|
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
algorithm.insectfiltering.enabled | 2 | 2 | Insect and outlier filtering enabled |
|
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
mor.limit.max | 2 | 2 | Sets the maximum reported MOR (in km). | 1-100 |
mor.scattercorr.method | 2 | 2 | Select correction for dust or disable correction |
|
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
data.message.mes2 | 0 | − | Message 2 | |
data.message.mes7 | 0 | − | Message 7 | |
data.message.mes8 | 0 | − | Message 8 | |
data.message.status | 0 | − | Status |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
fd.factory.calibrationdate | 0 | − | Factory calibration date | − |
fd.factory.sensitivitymainreceiver | 0 | − | Sensitivity of main receiver at factory | − |
fd.factory.sensitivitysidereceiver | 0 | − | Sensitivity of side receiver at factory | − |
fd.factory.offsetmainreceiver | 0 | − | Offset of main receiver at factory | − |
fd.factory.offsetsidereceiver | 0 | − | Offset of side receiver at factory | − |
fd.factory.cleanrefmainreceiver | 0 | − | Clean value of main receiver window at factory | − |
fd.factory.cleanrefsidereceiver | 0 | − | Clean window reference of side receiver at factory | − |
fd.factory.cleanreftransmitter | 0 | − | Clean value of transmitter window at factory | − |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
fd.field.calibrationdate | 0 | − | Latest field calibration date | |
fd.field.sensitivitymainreceiver | 0 | − | Sensitivity of main receiver | − |
fd.field.sensitivitysidereceiver | 0 | − | Sensitivity of side receiver | − |
fd.field.offsetmainreceiver | 0 | − | Offset of main receiver | − |
fd.field.offsetsidereceiver | 0 | − | Offset of side receiver | − |
fd.field.cleanrefmainreceiver | 0 | − | Clean value of main receiver window | − |
fd.field.cleanrefsidereceiver | 0 | − | Clean window reference of side receiver | − |
fd.field.cleanreftransmitter | 0 | − | Clean value of transmitter window | − |
fd.field.windowlifespanmainreceiver | 0 | − | Main receiver window lifespan | − |
fd.field.windowlifespansidereceiver | 0 | − | Side receiver window lifespan | − |
fd.field.windowlifespantransmitter | 0 | − | Transmitter window lifespan | − |
It is possible to adjust the level of window contamination (in percentage) required to produce an indication (I), warning (W), or alarm (A). You can adjust the limits separately for transmitter, receiver(s), and path(s). |
WARNING
Contact Vaisala before changing any warning or alarm limits. If you set the contamination alarm limit too high, measurement quality may become compromised without the device reporting about the condition. |
Parameter | GET | SET | Description | Value / Range |
---|---|---|---|---|
alerts.fd.operatingvoltagelow | 2 | − | Operating voltage lower limit | |
alerts.fd.operatingvoltagehigh | 2 | − | Operating voltage upper limit | |
alerts.fd.heatingvoltagelow | 2 | − | Heating voltage lower limit | |
alerts.fd.heatingvoltagehigh | 2 | − | Heating voltage upper limit | Default: 33.0 |
alerts.fd.batteryoperation | 2 | − | Battery operation detection limit | |
alerts.fd.mainpathcontaminationi | 2 | 2 | Main measurement path contamination indication limit | 0–100 Default: 15 |
alerts.fd.mainpathcontaminationw | 2 | 2 | Main measurement path contamination warning limit | 0–100 Default: 20 |
alerts.fd.mainpathcontaminationa | 2 | 2 | Main measurement path contamination alarm limit | 0–100 |
alerts.fd.invalidcleanwindowreferencew | 2 | − | Invalid clean window reference warning limit | -100 … 0 |
alerts.fd.invalidcleanwindowreferencea | 2 | − | Invalid clean window reference alarm limit | -100 … 0 |
alerts.fd.senscalfactlow | 2 | − | Receiver sensitivity low limit | 0.0–10.0 |
alerts.fd.senscalfacthigh | 2 | − | Receiver sensitivity high limit | 0.0–10.0 |
alerts.fd.offsetlow | 2 | − | Receiver offset low limit | -1.0 … 1.0 |
alerts.fd.offsethigh | 2 | − | Receiver offset high limit | -1.0 … 1.0 |
alerts.fd.cleanwindowreflow | 2 | − | Clean window reference low limit | 0.0–1000.0 |
alerts.fd.cleanwindowrefhigh | 2 | − | Clean window reference high limit | 0.0–1000.0 |
alerts.fd.contaminationcoefficientlow | 2 | − | Contamination fitting coefficient low limit | 0.0–1.0 |
alerts.fd.contaminationcoefficienthigh | 2 | − | Contamination fitting coefficient high limit | 0.0–1.0 |
alerts.fd.sidepathcontaminationi | 2 | 2 | Side measurement path contamination indication limit | 0–100 |
alerts.fd.sidepathcontaminationw | 2 | 2 | Side measurement path contamination warning limit | 0–100 |
alerts.fd.sidepathcontaminationa | 2 | 2 | Side measurement path contamination alarm limit | 0–100 |
alerts.fd.recentlystarted | 2 | − | Recently started indication limit | 0–3600 |
alerts.fsc.emmclifetimew | 2 | − | eMMC flash memory lifetime warning limit | 0–10 |
alerts.fsc.emmclifetimea | 2 | − | eMMC flash memory lifetime alarm limit | 0–10 |
alerts.fst.lightsourcecurrentlow | 2 | − | Transmitter light source current lower limit | Default 20.0 |
alerts.fst.lightsourcecurrenthigh | 2 | − | Transmitter light source current upper limit | Default 115.0 |
alerts.fst.lightsourceageindex | 2 | 2 | Limit over which light source age alert is raised | |
alerts.fst.opticalpowerlow | 2 | − | Transmitter optical power lower limit | |
alerts.fst.opticalpowerhigh | 2 | − | Transmitter optical power upper limit | |
alerts.fst.opticalpowerlastok | 2 | − | Transmitter optical power last good value | |
alerts.fst.windowcontaminationi | 2 | 2 | Transmitter window contamination indication limit | 0–100 Default: 15 |
alerts.fst.windowcontaminationw | 2 | 2 | Transmitter window contamination warning limit | 0–100 Default: 20 |
alerts.fst.windowcontaminationa | 2 | 2 | Transmitter window contamination alarm limit | 0–100 Default: 30 |
alerts.fst.windowblocked | 2 | 2 | Transmitter window blocked alarm limit | 0–100 |
alerts.fst.windowlifespanw | 2 | − | Transmitter window lifespan warning limit | 0–100 |
alerts.fst.windowlifespana | 2 | − | Transmitter window lifespan alarm limit | 0–100 |
alerts.fsrm.windowcontaminationi | 2 | 2 | Main receiver window contamination indication limit | 0–100 |
alerts.fsrm.windowcontaminationw | 2 | 2 | Main receiver window contamination warning limit | 0–100 |
alerts.fsrm.windowcontaminationa | 2 | 2 | Main receiver window contamination alarm limit | 0–100 |
alerts.fsrm.windowblocked | 2 | 2 | Main receiver window blocked alarm limit | 0–100 |
alerts.fsrm.ambientlightw | 2 | − | Main receiver ambient light warning limit | 0–10000 |
alerts.fsrm.ambientlighta | 2 | − | Main receiver ambient light alarm limit | 0–10000 |
alerts.fsrm.windowlifespanw | 2 | − | Main receiver window lifespan warning limit | 0–100 |
alerts.fsrm.windowlifespana | 2 | − | Main receiver window lifespan alarm limit | 0–100 |
alerts.fsrs.windowcontaminationi | 2 | 2 | Side receiver window contamination indication limit | 0–100 |
alerts.fsrs.windowcontaminationw | 2 | 2 | Side receiver window contamination warning limit | 0–100 |
alerts.fsrs.windowcontaminationa | 2 | 2 | Side receiver window contamination alarm limit | 0–100 |
alerts.fsrs.windowblocked | 2 | 2 | Side receiver window blocked alarm limit | 0–100 |
alerts.fsrs.ambientlightw | 2 | − | Side receiver ambient light warning limit | 0–10000 |
alerts.fsrs.ambientlighta | 2 | − | Side receiver ambient light alarm limit | 0–10000 |
alerts.fsrs.windowlifespanw | 2 | − | Side receiver window lifespan warning limit | 0–100 |
alerts.fsrs.windowlifespana | 2 | − | Side receiver window lifespan alarm limit | 0–100 |
When FD70 is installed in a high-humidity environment with strong condensation, the
standard heater settings may not be enough to keep FD70 windows free from condensation. In
high-humidity environments, Vaisala recommends using the following
settings.
|