Before you can give any commands to the
device, you must open a maintenance connection locally at the site. The maintenance
connection is an Ethernet connection.
Opening a maintenance connection through
an Ethernet connection does not interrupt measurement data transfer.
The local Ethernet maintenance connection
does not require an open command.
If you have 2 serial lines, you can
configure one port for data and the other for maintenance (CLI). In that case, there is
no need for the open command.
If you have only 1 serial line, use the
open command to change from the
DATA mode to the CLI mode. This
interrupts measurement data transfer.
Connect the Ethernet maintenance cable to FD70.
If you have the FD70 interface unit, connect the cable to the maintenance port.
CAUTION Remove the Ethernet maintenance cable after setup. It is not
intended for permanent installation.
If you have the measurement unit only, connect the cable to the open-ended cable.
Measurement unit cable pinout
Pin
mm2
Wire color
Description
7
1
Black
Operating voltage Vin−
6
1
Red
Operating voltage Vin+
5
1
White-red
Heating voltage VH+
8
1
White-black
Heating voltage VH−
13
0.14
Violet
COM3: RS-485A
2
0.14
White-violet
COM3: RS-485B
1
0.25
Brown
Day/Night switch
14
0.14
White-orange
Ethernet TX+
4
0.14
Orange
Ethernet TX−
15
0.14
White-green
Ethernet RX+
3
0.14
Green
Ethernet RX−
10
0.14
White-blue
COM1: GND1
16
0.14
Pink
COM1: RS-232 TX / RS-485B
9
0.14
White-pink
COM1: RS-232 RX / RS-485A
11
0.25
Blue
COM2: GND2
12
0.14
Gray
COM2: RS-232 TX / RS-485B
17
0.14
White-gray
COM2: RS-232 RX / RS-485A
Open the command prompt and type the following command:
Through the COM
port:
open <device type ID> <device ID>
For
example:
openFD1
If
you are opening the connection for the first time, there is no device ID.
If there is only one device on the bus,
no IDs are
needed.
open
The
port changes to the CLI mode.
Through the maintenance port as an
admin user:
ssh admin@<fixed IP address of FD70>
Type the password.
If there is no traffic, the port returns
from the CLI mode to the DATA mode after a
defined time. The default timeout period is 10 minutes.