Check that the enclosure grounding cable is connected to the system grounding point.
- 1
- Cable lug with hole Ø 6.2 mm (0.24 in)
- 2
- System grounding point
- 3
- AC (mains) cable
- 4
- Enclosure grounding cable
-
Check that the other end of the enclosure grounding cable is connected to a grounded location such as grounding connector.
- 1
- System grounding point
- 2
- Enclosure grounding cable
- 3
- Grounding connector
- 4
- Grounding cables
-
Check that the grounding cables have not been pulled too tight.
CAUTION Keep the curve of the down conductor smooth with a minimum radius of the bent cable 20 cm (8 in) and never over 90° angles.