Why:
-
Plate adjustment minimizes the effects caused by different installation heights and differences between individual mobile sensors. Repeat this procedure when changing the installation position.
How:
The reference setting is performed using the reference plate.
Which commands are needed to implement this?
Command: SET REFERENCES
- With parameter 0 (PLATE)
-
Action takes typically 30 ... 60 seconds
What the command does?
- Changes reference value parameters 0x50, 0x51, and 0x52 (1 parameter for each laser).
- When started, changes status bit 1 (reference setting ongoing) to 1 = ongoing.
- When finished, changes back to 0 = not ongoing.
What is needed for command to succeed?
- Sensor should not report errors related to reference setting.
How do I know did the command succeed?
- If status of reference setting ongoing is 0 = not ongoing.
- There are no errors.