Configuring measurement interval - AQT400

AQT400 Series Configuration Guide

Document code
M212059EN
Revision
C
Language
English
Product
AQT400
Document type
User guide
The measurements of the particle counter are based on the number of particles sampled over a period of 60 seconds, and the default measurement interval is 10 minutes.
  1. Establish a maintenance connection to AQT400. See Connecting to AQT400 with RS-232 (CLI).
  2. Set the measurement interval in minutes.

    Measurement interval defines how often AQT400 takes measurements.

    To define the measurement interval to be 3 minutes, type the following:

    set lpc_int=3
    write --really

    Measurement interval 3 ... 255 minutes, default 10 minutes.