Configuring Modbus address - AQT400

AQT400 Series Configuration Guide

Document code
M212059EN
Revision
C
Language
English
Product
AQT400
Document type
User guide
  1. Establish a connection to AQT400. See Connecting to AQT400 with RS-232 (CLI).
  2. To change the Modbus address and save the configuration change to the non-volatile memory, type the following:
    set rs485_addr=<address>
    write --really

    Modbus address range: 1 ... 247, default = 1.

  3. To reboot the device and check that the settings is correct, type the following:
    reboot --really
    show rs485_addr
  4. When done, close the connection.