The following parameters are available in AQ Tool for configuring communication and data logging.
Table 1. AQ Tool Configuration Parameters
Parameter Alias Description
GSM APN address gsm_apn Access point name (APN) supplied by the SIM card provider. Used in wireless communication with MOG100.
GSM username gsm_username Username supplied by the SIM card provider. Used in wireless communication with MOG100.
GSM password gsm_password Password supplied by the SIM card provider. Used in wireless communication with MOG100.
AQ-View address #1 gw_host1 Defines the address to which MOG100 sends the data. Used to configure use with Beacon Cloud, NM10, or a third-party server.
AQ-View address #2 gw_host2 Backup address for AQ-View address #1.
AQ-View address #3 gw_host3 Backup address for AQ-View address #2.
AQ-View enabled gw_enable

0 = wireless communication disabled

1 = wireless communication enabled

Automatic configuration autoconf

Defines whether to use automatic unit configuration. Enabled by default.

0 = disabled

1 = enabled

Unit configuration unitconf

Shows the devices that are connected to MOG100. The value is generated automatically.

0 = no devices

1 = AQT

2 = WXT

3 = AQT and WXT

Communication interval commint Defines how often measurement data is transferred from MOG100 to Beacon Cloud or another service. The default is 10 minutes.
Measurement interval measint

Defines the interval between individual measurements that are registered in MOG100.

The default is 5 minutes.

To set the measurement interval of AQT400, use Beacon Cloud.

Logging interval logint

Defines the interval in minutes for storing the measurement values in MOG100.

0 = logging disabled

To save the edited parameters in AQ Tool, select Write to eeprom.