This procedure tells you how to use the indicator to adjust the probe's measurement in 1 ... 2 points.
- Connect the probe(s) to the indicator.
-
Open the
main menu by pressing
.
- Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
-
If your probe supports the sensor purge feature, wait for the start-up purge to finish
or start the purge manually by selecting Devices > <your device> > Sensor purge.
After the purge has been completed, continue to the next step.
- Select Calibrate / Adjust.
- Select the parameter, for example, Relative humidity.
- Select Manual adjustment.
-
Place the probe to the first reference point and select Reference value, point 1 to open the stabilizing graph view. Wait for the measurement to stabilize fully.
The stabilization typically takes at least 10 ... 15 minutes.
-
When the reading has stabilized, press
.
- Use the arrow buttons to set the value of the reference, and then select OK or Set.
If you already have the reference and measured values available for your calibration point(s), you can just skip the stabilizing graph view with the button to get to the view where you can set the reference value. Then replace the automatically filled Measured value, point 1 and Measured temperature, point 1 with your actual values.
-
When the reading has stabilized, press
-
If you want to adjust the measurement in 2 points, place the probe to the second
reference point. Then scroll down the <parameter> adjustment
view and select Reference value, point 2 to open the stabilizing graph view. Wait for the measurement to stabilize fully.
-
When the reading has stabilized, press
.
- Use the arrow buttons to set the value of the second reference, and then select OK or Set.
-
When the reading has stabilized, press
-
In <parameter> adjustment view, check the difference
between each reference and measured value. Very large differences may be due to
insufficient stabilization time or unsuitable calibration setup.
- To store the adjustments to the probe, scroll down and select Activate adjustment > OK. Confirm by selecting OK.
- To exit without taking the
adjustment in use, press
to end calibration.
-
Update the calibration information in Calibrate / Adjust > Calibration information. Navigate to the Calibration information menu by pressing
.
-
Press
to end calibration.