Property | Description/Value |
---|---|
Water activity | |
Measurement range | 0 … 1 aw |
T90 response time 1 | 10 min |
Sensor | HUMICAP® 180L2 |
Accuracy 2 |
±0.01 aw (±1 %RS) |
Water concentration in oil | |
Typical accuracy | 10 % of the reading |
Temperature | |
Measurement range | −40 … +180 °C (−40 … +356 °F) |
Accuracy at +20 °C (+68 °F) | ±0.2 °C (0.36 °F) |
Property | Description/Value |
---|---|
Operating temperature of probe head | −40 … +180 °C (−40 … +356 °F) |
Operating temperature of probe body | −40 … +80 °C (−40 … +176 °F) |
Storage temperature range | −40 … +80 °C (−40 … +176 °F) |
Operating pressure range | 0 … 40 bar (0 … 580 psia) |
Installation pressure | Up to 10 bar (145 psia) |
IP rating of probe body |
IP66: Dust-tight. Protected from powerful water jets from any direction. |
Ball valve | |
Operating temperature | Up to +120 °C (+248 °F) |
Operating pressure | Up to 40 bar (0 … 580 psia) |
Property | Description/Value |
---|---|
Operating voltage | 15 … 30 V DC |
Current consumption | 10 mA typical |
Digital output | RS-485, non-isolated |
Protocols | Modbus RTU |
Output parameters |
Relative saturation (%RS) Temperature (°C) Water activity Water concentration in oil (ppmv) |
Property | Value/Description |
---|---|
EU directives and regulations |
EMC Directive (2014/30/EU) RoHS Directive (2011/65/EU) amended by 2015/863 |
Electromagnetic compatibility (EMC) |
EN 61326‑1, industrial environment |
Type approvals |
DNV GL certificate no. TAA00002YT |
Compliance marks |
CE, China RoHS, RCM |
Property | Description/Value |
---|---|
Connector | M12 5-pin A-coded male |
Weight | 510 g (18.0 oz) |
Filter options |
Stainless steel grid standard filter Stainless steel grid filter for high flow rates (> 1 m/s) |
Probe cable length | 2 m (6.56 ft) |
Adjustable installation depth | 35 … 179 mm (1.37 … 7.05 in) |
Materials | |
Probe | AISI 316L |
Probe body | AISI 316L |
Cable jacket | FEP |
1 At +20 °C (+68
°F) in still oil.
2 In range 0 … 0.5
aw, including non-linearity, hysteresis, and repeatability. See
accuracy graph below.