Configuring Minimum and Maximum RH and T Errors with Insight - SWINSIGHT10 - TMD62 - HMD62 - HMD60

HMD62 TMD62 User Guide

Document code
M212016EN
Revision
C
Language
English (United States)
Product
TMD62
HMD62
HMD60
Document type
User guide

You can define the minimum and maximum limit that the RH (HMD62 only) or T measurement output can reach before the output moves to error state.

Note that configuring these settings requires using Insight in Advanced Mode.
Figure 1. Min/Max Allowed Measurement Value Before Error (HMD62 Example)
To define the minimum/maximum error limits:
  1. Switch to Advanced Mode in the Settings menu.
  2. Select > Configure Device > Advanced.
  3. Enter the values for minimum and maximum RH and T as applicable, select Save and exit the menu with Close.