The status register (E000hex) can
report that one of the following notification types is present:
- 0
- No issues
- 1
- Info
- 2
- Warning
- 3
- Error
Additional details about the status reported in the register E000hex can be provided in the 16-bit register E001hex, as described in the following table.
Bit in E001hex | Meaning | Notes |
---|---|---|
0 (0000hex) | No errors | |
1 (0001hex) | WLAN on | Wireless configuration interface is activated and can be connected to. |
2 (0002hex) | Probe error | Connect to Indigo's wireless configuration interface and review the error information in the Status menu. Check display messages. |
4 (0004hex) | Low supply voltage | The supply voltage is not within the specified range (15 ... 30 VDC (20 ... 22 VAC) |
8 (0008hex) | WLAN error | See the WLAN troubleshooting instructions. |
16 (0010hex) | No measurement probe | Probe not connected or Indigo cannot detect the probe. |
32 (0020hex) | Probe warning | Connect to Indigo's wireless configuration interface and review the error information in the Status menu. Check display messages. |