Table 1. Inputs and Outputs
Property Specification
Digital output RS-485 Modbus RTU
Relays 2 configurable relays (VAC/VDC)
Device maximum specification (resistive load):
  • Max. switching power 30 W / 37.5 VA
UL-rated maximum specification (resistive load):
  • AC: max. 28 V / 0.5 A
  • DC: max. 40 V / 0.24 A
  • Up to 30 VDC:
    • max. switching current 1 A
    • max. switching power 30 W

Power supply input 1

15 ... 30 VDC (20 ... 22 VAC)

Maximum current Transmitter and connected probe max. 1 A
Power consumption Transmitter max. 3 W (+ connected probe, varies depending on probe type)
Probe connector M12/5 connector for probe or probe cable connection (Vaisala Indigo-compatible probes)
Cable feed throughs 2 options: rubber lead-through on the bottom of the transmitter, and opening with a seal at the back of the transmitter
Screw terminal wire size 0.2 ... 1.5 mm2
CAUTION Do not modify the unit or use it in ways not described in the documentation. Improper modification may lead to safety hazards, equipment damage, failure to perform according to specification, or decreased equipment lifetime.
1 Using a power supply with overload protection is recommended for electrical safety.