In order to be able to use the calibration options, you must switch the operation of the probe and Indigo to calibration mode with the Start calibration button.
Figure 1. Start Calibration Button
1
Start calibration button on the Calibration tab

When you start the calibration mode, the Start calibration button is replaced with the Stop calibration button. The calibration mode remains active until you close it by selecting Stop calibration.

You can use other menus while the calibration mode is active, and return to the Calibration menu later to complete your adjustments.

Always close the calibration mode to return the probe and Indigo to normal operating mode. The measurement performance of the probe can be affected when used in calibration mode. You must close the calibration mode with the Stop calibration button also when no changes are made.