Syntax | Description |
---|---|
rsel [channel]<cr> | Show current relay settings. channel = Analog output channel |
rsel [channel parameter release_ppm set_ppm release_value set_value startup_value error_value]<cr> | Configure relay operation of the channel. channel = Analog output channel
parameter = The parameter that is output on analog channel. The only option is CO2 (in ppm). release_ppm = CO2 level at which the relay is released. set_ppm = CO2 level at which the relay is activated. release_value = The current output (mA) when the relay is released. set_value = Current (mA) output when the relay is activated. Set to 12 when using the relay box with GMP251/252. startup_value = During probe start-up, if the CO2 ppm level falls between the release_ppm and set_ppm values, startup_value shows the state of the relay (mA): the relay is either activated, released, or in error state. error_value = The current (mA) that is output when the probe is in an error state. |
Syntax | Description |
---|---|
rsel [channel]<cr> | Show current relay settings. channel = Analog output channel |
rsel [channel parameter release_ppm set_ppm set_value error_value]<cr> | Configure relay operation of the channel. channel = Analog output channel
parameter = The parameter that is output on analog channel. The only option is CO2 (in ppm). release_ppm = CO2 level at which the relay is released. set_ppm = CO2 level at which the relay is activated. set_value = Current (mA) output when the relay is activated. Set to 12 when using the relay box with GMP251/252. error_value = The current (mA) that is output when the probe is in an error state. |
The relay mode of GMP251 and GMP252 probes has been designed to use 12 mA as the relay control current. Use only 12 mA for the set_value configuration when using the relay box with GMP251 and GMP252. |