The configuration options in Indigo
transmitters include probe-specific settings that vary depending on the features of the
connected probe.
In the case of Vaisala CARBOCAP® Carbon Dioxide Probes GMP251 and GMP252, the probe-specific
settings allow configuring the filtering factor and environmental compensations.
To change the environmental compensation
settings:
-
Launch the Indigo wireless configuration interface and open the menu.
-
Enable the required environmental compensations (set the compensation
On or Off) you need to use in the
Measurement selections. For temperature compensation, select
either the measurement from the probe sensor (Measured), or manual
Setpoint entry.
-
Enter the setpoint values for the compensations you have enabled into the
Compensation setpoints text fields.
| The
values entered into the Compensation setpoints fields are only
for temporary use: when you reset the probe, the default values (see
Compensation power-up defaults) are restored. |

-
Enter the power-up default compensation values that are stored into the probe's
permanent memory.

| The values entered into the Compensation power-up defaults
fields remain in use also after probe reset. |
-
The values you enter are saved automatically when you exit the input field (for
example, tap on an area outside of the field).
| Note that the environmental
compensations you set in the and the compensations set on the Configuration tab of
the Calibration menu are interconnected: the configuration set in
either menu is applied to both. |