| Pin # | Function | Note | Wire colors 1 |
|---|---|---|---|
| 1 | Power in |
Digital output: 12–30 V DC Voltage output: 12–30 V DC Current output: 20–30 V DC Average consumption < 0.4 W, maximum 0.5 W |
|
| 2 |
RS-485- or voltage output |
Voltage output: 0–5 V DC or 0–10 V DC 2 |
|
| 3 | GND | ||
| 4 |
RS-485+ or current output |
Current output: 0–20 mA or 4–20 mA 2 |
|
| 5 | Output control |
Connecting pin #5 to GND (pin #3) forces the probe to analog output mode. If an analog output configuration has not been selected, default 0–10 V DC and 4–20 mA scalings are used. If pin #5 is not connected, the analog or digital output selected when ordering or set later through configuration is used. |
Replacing the filter
To open the filter, turn the filter counter-clockwise.Reinstall the filter to finger tightness.
Caution: Do not open the probe body. Be careful not to touch any parts under the filter.