Probe compatibility - HPP271

HPP271 User Guide

Document code
M211888EN
Revision
E
Language
English
Product
HPP271
Document type
User guide

The Indigo80 handheld indicator is tested for compatibility with HPP270 probes that have firmware version 1.4.0 or newer. Probes with older firmware versions may have limited compatibility with the indicator.

CAUTION

Do not use Indigo80 in low-power mode if you are logging data in a hydrogen peroxide environment. Using the low-power mode can power off HPP270 probes with a firmware version earlier than 1.4.0, resulting in condensation on the sensor and irreparable sensor damage.

In all hydrogen peroxide measurements, ensure that the automatic power off function is disabled and there is enough battery charge to avoid an accidental power-off. Using an AC adapter to power Indigo80 during long-term measurements is highly recommended.

For the most up-to-date version compatibility information, see Indigo80-compatible Firmware Technical Note (M212901EN).