HPP271
| Property | Description/Value |
|---|---|
| Hydrogen peroxide | |
| Sensor | PEROXCAP® |
|
Measurement range |
0 … 2000 ppm |
| Measurement temperature range | +5 … +50 °C (+41 … +122 °F) |
| Repeatability at +25 °C (+77 °F): | |
| up to 500 ppm H2O2 | ±10 ppm |
| up to 1000 ppm H2O2 | ±15 ppm |
| Accuracy at +10 … +25 °C (+50 … +77 °F) , 10 … 2000 ppm H2O2 1 | ±10 ppm or 5 % of reading (whichever is greater) |
| Factory calibration uncertainty at +25 °C (+77 °F), 500 ppm H2O2 2 | ±10 ppm |
| Response time (T63) | 70 s |
| Other parameters | |
| H2O ppm by volume, dew point temperature | |
| Property | Description/Value |
|---|---|
| Operating voltage |
Digital output: 15 … 30 V DC Analog output: 15 … 25 V DC |
| Current consumption at +25 °C (+77 °F) | |
| In digital mode | Max. 10 mA |
| In analog mode | Max. 50 mA |
| During sensor purge | Max. 250 mA |
| Digital output | |
| Interface | RS-485, not isolated; do not use termination on the RS-485 line |
| Communication protocol | Modbus RTU v.1.02 |
| Analog output | |
| Outputs | 2 × 4 … 20 mA 3-wire current outputs |
| Max. load | 500 Ω |
| Accuracy (typical) | ±0.1 % of full scale |
| Analog output temperature dependence | 0.005 %/°C (0.003 %/°F) full scale |
| Property | Description/Value |
|---|---|
| Operating temperature | +0 … +70 °C (+32 … +158 °F) |
| Storage temperature | −20 … +70 °C (−4 … +158 °F) |
| Ambient pressure | Normal atmospheric pressure |
|
IP rating |
IP65: Dust-tight. Protected from water jets from any direction. |
HPP272
| Property | Description/Value |
|---|---|
| Hydrogen peroxide | |
| Sensor | PEROXCAP® |
|
Measurement range |
0 … 2000 ppm |
| Measurement temperature range | +5 … +50 °C (+41 … +122 °F) |
| Repeatability at +25 °C (+77 °F): | |
| up to 500 ppm H2O2 | ±10 ppm |
| up to 1000 ppm H2O2 | ±15 ppm |
| Accuracy at +10 … +25 °C (+50 … +77 °F) , 10 … 2000 ppm H2O2 3 | ±10 ppm or 5 % of reading (whichever is greater) |
| Factory calibration uncertainty at +25 °C (+77 °F), 500 ppm H2O2 4 | ±10 ppm |
| Response time (T63) | 70 s |
| Relative saturation | |
| Measurement range | 0 … 100 %RS |
| Measurement temperature range | +5 … +50 °C (+41 … +122 °F) |
| Repeatability at +25 °C (+77 °F), 500 ppm H2O2 | ±0.5 %RS |
| Accuracy at +25 °C (+77 °F) 3 | ±4 %RS |
| Factory calibration uncertainty at +25 °C (+77 °F), 500 ppm H2O2 4 | ±2 %RS |
| Relative humidity | |
| Measurement range | 0 … 100 %RH |
| Measurement temperature range | +5 … +70 °C (+41 … +158 °F) |
| Accuracy: 3 | |
| at +25 °C (77 °F), 0 ppm H2O2, 0 … 90 %RH | ±1 %RH |
| over full temperature measurement and H2O2 range | ±2 %RH |
| Response time (T63) | 20 s |
| Factory calibration uncertainty at +25 °C (77 °F), 0 ppm H2O2: 4 | |
| at 0 … 40 %RH | ±0.6 %RH |
| at 40 … 95 %RH | ±1 %RH |
| Temperature | |
| Sensor | Pt1000 RTD Class F0.1 |
| Accuracy over temperature range | ±0.2 °C (±0.36 °F) |
| Other parameters | |
| Absolute H2O2 and H2O, H2O ppm by volume, water vapor saturation pressure (H2O and H2O+H2O2), dew point temperature, vapor pressure (H2O and H2O2) | |
| Property | Description/Value |
|---|---|
| Operating voltage |
Digital output: 15 … 30 V DC Analog output: 15 … 25 V DC |
| Current consumption at +25 °C (+77 °F) | |
| In digital mode | Max. 10 mA |
| In analog mode | Max. 50 mA |
| During sensor purge | Max. 250 mA |
| Digital output | |
| Interface | RS-485, not isolated; do not use termination on the RS-485 line |
| Communication protocol | Modbus RTU v.1.02 |
| Analog output | |
| Outputs | 2 × 4 … 20 mA 3-wire current outputs |
| Max. load | 500 Ω |
| Accuracy (typical) | ±0.1 % of full scale |
| Analog output temperature dependence | 0.005 %/˚C (0.003 %/°F) full scale |
| Property | Description/Value |
|---|---|
| Weight | 130 g (4.58 oz) |
| Connector | M12-5M |
| Probe body material | AISI316L stainless steel |
| Filter cap material | Porous PTFE |
| Temperature probe material | AISI316L stainless steel |
| Temperature probe cable material | PTFE |
| Property | Description/Value |
|---|---|
| Operating temperature | +0 … +70 °C (+32 … +158 °F) |
| Storage temperature | −20 … +70 °C (−4 … +158 °F) |
| Ambient pressure | Normal atmospheric pressure |
|
IP rating |
IP65: Dust-tight. Protected from water jets from any direction. |
1 Including non-linearity, hysteresis, and repeatability.
2 Defined as ±2 standard deviation limits. See also calibration certificate.
3 Including non-linearity, hysteresis, and repeatability.
4 Defined as ±2 standard
deviation limits. See also calibration certificate.