Status registers - OPT100

OPT100 User Guide

Document code
M211858EN
Revision
N
Language
English (United States)
Product
OPT100
Document type
User guide
Table 1. Status registers (read-only)

Only the first bit of each status bit field is used.

Register number Address Register description Data format Notes
1 0000hex Device stopped 16-bit field

0 = Device measuring normally

1 = Device stopped (either automatically or by administrator), being initialized, or being uninstalled

3 0002hex Measurement reliability 16-bit field

0 = Measurement is unreliable because of high ambient temperature

1 = Measurement is reliable

4 0003hex Gas alerts muted 16-bit field

0 = Gas level alerts operating normally

1 = Gas level alerts muted