Green |
DGA monitor operates normally and performs measurement cycles. |
Green (flashing) |
DGA monitor is in standby mode, or running a service tool. Requires user
interaction to begin measurement. |
Yellow |
DGA monitor requires user interaction. For example, a device warning is active,
a service tool requires user input, or the DGA monitor is being initialized or
uninstalled. |
Yellow (flashing) |
Gas level alert is active, caution limit exceeded. |
Red |
Device error is active. User interaction is required. |
Red (flashing) |
Gas level alert is active, alarm limit exceeded. |