Verifying safety after repair - OPT100
OPT100 User Guide
- Document code
- M211858EN
- Revision
- AA
- ft:locale
- en-US
- Product
- OPT100
- Document type
- User guide
Go through the following verification
steps after performing repairs on the DGA monitor.
-
If you have replaced or altered any mechanical parts of the DGA monitor cabinet or
mounting accessories, verify that:
- Cabinet can be opened and closed
normally.
- Cabinet tightness is maintained.
- All attachment points are
intact.
-
If you have replaced any electrical components or wiring, verify that:
- All components that are secured to a DIN rail are firmly in place.
- There are no disconnected cables or loose wires.
- All grounding wires are attached.
- Wiring is done according to the
wiring instructions in the OPT100 Installation Guide
(M211857EN), unless intentionally altered by the repair procedure.
-
If you have replaced any of the oil handling components (pipes, valves, pump, oil
block, etc.) or opened any of the oil connections, verify that:
- There are no open oil connections.
- All oil handling components are securely attached.
- All oil leaks have been cleaned and wiped clean. No oil is visible on the parts or
oil connections.