CAUTION If the Vaisala-supplied instrument power cable is not available, use a shielded cable with 6 mm2 (10 AWG) wires. Maximum length of the cable is 1.2 m (3 ft 11 in). |
- Open the cable gland marked DC in on the DGA monitor. Remove the plug and store it for later use.
- Insert the instrument power cable through the outer nut and the seal insert, and start inserting the cable through the cable gland. Stop when you reach the section of the cable where the metal braid is exposed.
- Bend the cable braid over the seal insert so that it will make contact with metal when the cable gland is tightened.
- Push the outer nut and the seal insert against the contact socket of the gland and tighten the outer nut.
-
Inside the DGA monitor, connect the wires to terminal block Y3:
Table 1. Terminal block Y3 wiring Signal Wire color
Vaisala cable CBL210544
Terminal 24 V DC + Black with red marking + 24 V DC − Black − Ground Green/yellow - Open the cable gland marked DC out on the power supply unit. Remove the plug and store it for later use.
- Insert the instrument power cable through the outer nut and the seal insert, and start inserting the cable through the cable gland. Stop when you reach the section of the cable where the metal braid is exposed.
-
Bend the cable braid over the seal insert so that it will make contact with metal when
the cable gland is tightened.
- Push the outer nut and the seal insert against the contact socket of the gland and tighten the outer nut.
-
Inside the power supply unit, connect the wires to terminal block X5:
Table 2. Terminal block X5 wiring Signal Wire color
Vaisala cable CBL210544
Terminal 24 V DC + Black with red marking + 24 V DC − Black − Ground Green/yellow