OPTPSU cable glands and connectors - OPT100

OPT100 Installation Guide

Document code
M211857EN
Revision
U
Language
English
Product
OPT100
Document type
User guide
Figure 1. Cable glands and connectors on OPTPSU power supply unit
# Label Description Cable diameter Wrench size
1 Ground connection. 4 … 16 mm2 (12 … 5 AWG) conductor 13 mm
2 AC in (OPTPSU1) Mains power input. 100 … 240 V AC, 50 … 60 Hz, 10 A. 9 … 16 mm (0.35 … 0.63 in) 0.5 … 10mm2 (24 … 6 AWG) conductor 30 mm
DC in (OPTPSU2) Mains power input. 110 … 220 V DC, 6.3 … 2.5 A. 9 … 16 mm (0.35 … 0.63 in) 0.5 … 10mm2 (24 … 6 AWG) conductor 30 mm
3 Spare Not reserved. 5 … 10 mm (0.20 … 0.39 in) 20 mm
4 Spare Not reserved. 5 … 10 mm (0.20 … 0.39 in) 20 mm
5 Relay out Relay output, max 250 V AC, 10 A. 7 … 12 mm (0.28 … 0.47 in) 24 mm
6 Relay control in Relay control from DGA monitor. Use Vaisala cable CBL210539. 7 … 12 mm (0.28 … 0.47 in) 24 mm
7 DC out DC output to DGA monitor, 24 V DC, 20 A. Use Vaisala cable CBL210544. 9 … 16 mm (0.35 … 0.63 in) 30 mm