0WF!0WF,g=A,l=N,t=A,0=273.00,1=273.01,2=273.00,3=273.00,4=273.00,5=273.00,a=45.1,b=50.2,u=54.9,v=63.1,x=65.1,y=65.1<cr><l f>
| Field character | Field name | Description |
|---|---|---|
| g | Strategy |
A = All N = North E = East S = South |
| l | Pulse length |
N = Normal, auto A = Adjusted on half S = Short E = Extended T = Test |
| t | Single transducer mode |
A = All N = North E = East S = South |
| 0..5 | Zero adjustment | 1 … 655.35 us (default 273.00 us) |
| a, b | Detect level btw. N and E | 0 … 100 % (default 70 %) |
| u, v | Detect level btw. E and S | 0 … 100 % (default 70 %) |
| x, y | Detect level btw. S and N | 0 … 100 % (default 70 %) |