| WXT530 must be in the CLI mode before the
firmware update. CLI is the command line interface, such as ASCII, NMEA, and SDI-12.
Firmware update is not possible in the Modbus mode. |
-
Connect the WXT530 service cable to the USB port of your computer and to the WXT530
service connector.
-
To make sure that WXT530 is in the CLI mode, open the terminal software, such as Tera
Term, with settings 19200, 8, N, 1 and type the command ?!.
If the sensor is in the CLI mode, the terminal program prints the sensor address.
-
Start WxtLoader.exe.
-
To change the serial port settings, select Configuration.
- Select the COM port in use.
- Select serial port settings: 19200, 8
bit, none, 1 bit.
-
When Status is Connected to device, select Upload.
When Upload
status is 100 %, the firmware update is completed.