Changing Modbus address in CLI mode - WXT530

WXT530 Series User Guide

Product status
Document code
M211840EN
Revision
G
ft:locale
en-US
Product
WXT530
Document type
User guide
You can use Vaisala Configuration Tool (version 2.62 or newer) to change between Modbus and ASCII protocol. Alternatively use a terminal software as follows.
  1. To change the communication settings, change to the CLI mode.
  2. To print the Modbus address, type ?XU,O! in the terminal software.
  3. Change the Modbus address.
    ?XU,O=1!
  4. To change back to the Modbus mode, see Changing from CLI mode to Modbus mode.