Start concurrent measurement (aC) - WXT530

WXT530 Series User Guide

Product status
Document code
M211840EN
Revision
G
ft:locale
en-US
Product
WXT530
Document type
User guide

Use this command when there are several devices on the same bus and simultaneous measurements are needed from the devices, or if more than 9 measurement parameters are requested from a single device.

The measured data is not sent automatically. You must request it with the Send data command aD.

Command format:

aCx!
a Device address
C Start concurrent measurement command
x The desired measurement
1 = Wind
2 = Temperature, humidity, and pressure
3 = Precipitation
4 = Analog input
5 = Supervisor
If x is left out, the query refers to combined data message in which the user can request data from several sensors with just one command. See the following example.
! Command terminator

Response:

atttnn<cr><lf>
a Device address
ttt The measurement completing time in seconds
nn The number of the measured parameters available (maximum number is 20)
<cr><lf> Response terminator
For information on changing the message parameters, units and other settings, see Sensor and data message settings.