Use this command to check the current precipitation sensor settings.

Command format in ASCII and NMEA 0183:

aRU<cr><lf>

Command format in SDI-12:

aXRU!
a Device address
RU Precipitation sensor settings command in ASCII and NMEA 0183
XRU Precipitation sensor settings command in SDI-12
<cr><lf> Command terminator in ASCII and NMEA 0183
! Command terminator in SDI-12

The response in ASCII and NMEA 0183:

aRU,R=[R],I=[I],U=[U],S=[S],M=[M],Z=[Z],X=[X],Y=[Y]<cr> <lf>

The response in SDI-12:

aXRU,R=[R],I=[I],U=[U],S=[S],M=[M],Z=[Z],X=[X],Y=[Y]<cr><lf>

where [R][I][U][S][M][Z][X][Y] are the setting fields.

Example (ASCII and NMEA 0183, device address 0):

0RU<cr><lf>

0RU,R=11111100&10000000,I=60,U=M,S=M,M=R,Z=M,X=100, Y=100<cr><lf>

Example (SDI-12, device address 0):

0RU!0RU,R=11111100&10000000,I=60,U=M,S=M,M=R, Z=M,X=100,Y=100<cr><lf>