Aux input averaging time [A] - WXT530

WXT530 Series User Guide

Document code
M211840EN
Revision
F
Language
English (United States)
Product
WXT530
Document type
User guide

The Aux input averaging time in seconds. This parameter defines the averaging time for ultrasonic level sensor and solar radiation measurement. Pt1000 and the tipping bucket are not affected by this setting.

Make sure the averaging time is smaller than the update interval [I]. The smallest value 0.25 s means single measurement. A longer averaging time can decrease noise. A shorter averaging time gives a slightly smaller current consumption.