When changing the analog output setting, the setting is applied after the transmitter reset. When analog output is in use, the serial port does not work but the service connector does.

The 12th bit from left determines if analog output is enabled:

  • Analog output enabled at the factory: 0XF,f=1111111111110000
  • Analog output disabled at the factory: 0XF,f=1111111111100000

You can enable or disable analog output with the aWU,R= command.

  • To enable analog output: Change bit 7 to 1. 0WU,R=1111111111111100
  • To disable analog output: Change bit 7 to 0. 0WU,R=1111110111111100