CO2 sampling from incubator using pipe
1
Condensation window
2
Incubator wall
3
GMP80P sampling pipe inserted through the incubator sampling port
4
Connection cable to Indigo80
  • Recommended maximum sampling time with the pipe is 15 minutes. Monitor the level of condensation through the condensation window on top of GMP80P.
  • Let GMP80P dry between measurements. For example, after a sampling time of 5 minutes, 10 minutes of drying time is required.

To remove condensed water from GMP80P:

  1. Detach GMP80P from the incubator wall (do not disconnect from Indigo80).
  2. Place GMP80P on a flat surface and leave it running, making sure it does not fall off the surface.

    You can continue measuring when there is no more condensation visible in the window on top of GMP80P.