If needed, you can restore the factory
adjustment for your probe's output parameters using the Indigo80 handheld indicator.
|
Always restore factory adjustment before entering a new
adjustment. This prevents any possible earlier adjustments having an effect on the new
adjustment you make.
|
-
Connect the probe(s) to the indicator.
-
Open the indicator main menu by pressing .
-
Select Devices. If you have more than one device connected to the indicator, make a further selection
between the devices.
-
Select Calibrate / Adjust.
-
Select the type of adjustment to restore.
For example, select CO₂ adjustment.
-
Select Manual adjustment.
-
Select Enter manually.
-
Scroll down and select .
-
Confirm by selecting OK.
-
Press
to end calibration.