Pin# | Function | Note | Cable 223263SP wire colors |
---|---|---|---|
1 | Power in |
|
Brown |
2 | RS-485- or voltage output |
Voltage: 0–5 V DC or 0–10 V DC (default analog output scaling) |
White |
3 | GND | – | Blue |
4 | RS-485 + or current output |
Current: 0–20 mA or 4–20 mA (default analog output scaling) |
Black |
5 | Output control |
Connecting pin #5 to GND (pin #3) forces the probe to analog output mode. If an analog output configuration has not been selected, default 0–10 V DC and 4–20 mA scalings are used. If pin #5 is not connected, the analog or digital output selected when ordering or set later through configuration is used. |
Gray |
Note that the probe always remains in analog mode when pin #5 is connected to pin #3, and cannot be switched to digital output in this wiring option. |