Exception responses - GMP251 - GMP80P

GMP251 and GMP80P User Guide

Document code
M211799EN
Revision
K
Language
English
Product
GMP251
GMP80P
Document type
User guide
Modbus exception responses
Code Name Reason
01 ILLEGAL FUNCTION Unsupported function code
02 ILLEGAL DATA ADDRESS Register address or Object ID out of valid ranges
03 ILLEGAL DATA VALUE Otherwise invalid request

Accessing unavailable (temporarily missing) measurement data does not generate a Modbus exception. “Unavailable” value (a quiet NaN for floating point data or 0x8000 for integer data) is returned instead. An exception is generated only for any access outside the applicable register ranges.